Bioanalysis ; 9(7): 553-563, 2017 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-28300419

RESUMO

AIM: Hematocrit (Ht) effects remain a challenge in dried blood spot (DBS) sampling. The aim was to develop an immunosuppressant DBS assay on two LC-MS/MS systems covering a clinically relevant Ht range without Ht correction. RESULTS: The method was partially validated for tacrolimus, sirolimus, everolimus, cyclosporin A and fully validated for mycophenolic acid on an Agilent and Thermo LC-MS/MS system. Bias caused by Ht effects were within 15% for all immunosuppressants between Ht levels of 0.23 and 0.48 l/l. Clinical validation of DBS versus whole blood samples for tacrolimus and cyclosporin A showed no differences between the two matrices. CONCLUSION: A multiple immunosuppressant DBS method without Ht correction, has been validated, including a clinical validation for tacrolimus and cyclosporin A, making this procedure suitable for home sampling.


Assuntos
Cromatografia Líquida/métodos , Teste em Amostras de Sangue Seco/métodos , Monitoramento de Medicamentos/métodos , Imunossupressores/sangue , Espectrometria de Massas em Tandem/métodos , Ciclosporina/sangue , Everolimo/sangue , Hematócrito , Humanos , Ácido Micofenólico/sangue , Sirolimo/sangue , Tacrolimo/sangue