Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 10 de 10
Filtrar
1.
BMC Anesthesiol ; 17(1): 48, 2017 03 21.
Artigo em Inglês | MEDLINE | ID: mdl-28327093

RESUMO

BACKGROUND: Despite increased interest in non-invasive arterial pressure monitoring, the majority of commercially available technologies have failed to satisfy the limits established for the validation of automatic arterial pressure monitoring by the Association for the Advancement of Medical Instrumentation (AAMI). According to the ANSI/AAMI/ISO 81060-2:2013 standards, the group-average accuracy and precision are defined as acceptable if bias is not greater than 5 mmHg and standard deviation is not greater than 8 mmHg. In this study, these standards are used to evaluate the CareTaker® (CT) device, a device measuring continuous non-invasive blood pressure via a pulse contour algorithm called Pulse Decomposition Analysis. METHODS: A convenience sample of 24 patients scheduled for major abdominal surgery were consented to participate in this IRB approved pilot study. Each patient was monitored with a radial arterial catheter and CT using a finger cuff applied to the contralateral thumb. Hemodynamic variables were measured and analyzed from both devices for the first thirty minutes of the surgical procedure including the induction of anesthesia. The mean arterial pressure (MAP), systolic and diastolic blood pressures continuously collected from the arterial catheter and CT were compared. Pearson correlation coefficients were calculated between arterial catheter and CT blood pressure measurements, a Bland-Altman analysis, and polar and 4Q plots were created. RESULTS: The correlation of systolic, diastolic, and mean arterial pressures were 0.92, 0.86, 0.91, respectively (p < 0.0001 for all the comparisons). The Bland-Altman comparison yielded a bias (as measured by overall mean difference) of -0.57, -2.52, 1.01 mmHg for systolic, diastolic, and mean arterial pressures, respectively with a standard deviation of 7.34, 6.47, 5.33 mmHg for systolic, diastolic, and mean arterial pressures, respectively (p < 0.001 for all comparisons). The polar plot indicates little bias between the two methods (90%/95% CI at 31.5°/52°, respectively, overall bias = 1.5°) with only a small percentage of points outside these lines. The 4Q plot indicates good concordance and no bias between the methods. CONCLUSIONS: In this study, blood pressure measured using the non-invasive CT device was shown to correlate well with the arterial catheter measurements. Larger studies are needed to confirm these results in more varied settings. Most patients exhibited very good agreement between methods. Results were well within the limits established for the validation of automatic arterial pressure monitoring by the AAMI.


Assuntos
Pressão Arterial/fisiologia , Determinação da Pressão Arterial/instrumentação , Determinação da Pressão Arterial/métodos , Monitorização Intraoperatória/métodos , Idoso , Idoso de 80 Anos ou mais , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Projetos Piloto
2.
BMC Emerg Med ; 10: 1, 2010 Jan 15.
Artigo em Inglês | MEDLINE | ID: mdl-20078862

RESUMO

BACKGROUND: Although congestive heart failure (CHF) patients typically present with abnormal auscultatory findings on lung examination, respiratory sounds are not normally subjected to additional analysis. The aim of this pilot study was to examine respiratory sound patterns of CHF patients using acoustic-based imaging technology. Lung vibration energy was examined during acute exacerbation and after clinical improvement. METHODS: Respiratory sounds throughout the respiratory cycle were captured using an acoustic-based imaging technique. Twenty-three consecutive CHF patients were imaged at the time of presentation to the emergency department and after clinical improvement. Digital images were created (a larger image represents more homogeneously distributed vibration energy of respiratory sound). Geographical area of the images and respiratory sound patterns were quantitatively analyzed. Data from the CHF patients were also compared to healthy volunteers. RESULTS: The median (interquartile range) geographical areas of the vibration energy image of acute CHF patients without and with radiographically evident pulmonary edema were 66.9 (9.0) and 64.1(9.0) kilo-pixels, respectively (p < 0.05). After clinical improvement, the geographical area of the vibration energy image of CHF patients without and with radiographically evident pulmonary edema were increased by 18 +/- 15% (p < 0.05) and 25 +/- 16% (p < 0.05), respectively. CONCLUSIONS: With clinical improvement of acute CHF exacerbations, there was more homogenous distribution of lung vibration energy, as demonstrated by the increased geographical area of the vibration energy image.


Assuntos
Auscultação/métodos , Insuficiência Cardíaca/fisiopatologia , Sons Respiratórios/fisiopatologia , Adolescente , Adulto , Auscultação/instrumentação , Estudos de Casos e Controles , Diagnóstico por Computador/métodos , Ecocardiografia , Serviço Hospitalar de Emergência , Feminino , Insuficiência Cardíaca/complicações , Insuficiência Cardíaca/diagnóstico por imagem , Insuficiência Cardíaca/terapia , Humanos , Masculino , Pessoa de Meia-Idade , Projetos Piloto , Edema Pulmonar/complicações , Radiografia , Respiração Artificial , Processamento de Sinais Assistido por Computador , Espectrografia do Som/métodos , Resultado do Tratamento , Vibração , Adulto Jovem
3.
Respiration ; 77(2): 134-8, 2009.
Artigo em Inglês | MEDLINE | ID: mdl-19033680

RESUMO

BACKGROUND: Dyspnea is prevalent and has a broad differential diagnosis. Difficulty in determining the correct etiology can delay proper treatment. Non-invasively obtained acoustic signals may offer benefit in identifying patients with dyspnea due to obstructive airway disease (OAD). OBJECTIVES: The aim of this pilot study was to determine whether patients with acute dyspnea due to OAD had distinguishing features when studied with a computerized acoustic-based imaging technique. METHODS: Respiratory sounds from patients with dyspnea due to OAD (n = 32) and those with dyspnea not due to OAD (n = 39) were studied and compared with normal controls (n = 16). RESULTS: In patients without OAD and in controls, the ratios of peak inspiratory to peak expiratory vibration energy values (peak I/E vibration ratio) were remarkably similar, 6.3 +/- 5.1 and 5.6 +/- 4, respectively. For the OAD patients, the peak I/E vibration ratio was significantly lower at 1.3 +/- 0.04 (p < 0.01). In the patients without OAD and the controls, the ratios of inspiratory time to expiratory time (I/E time ratio) were again similar, 1.0 +/- 0.1 and 0.99 +/- 0.11, respectively. For the OAD patients, the I/E time ratio was significantly lower at 0.72 +/- 0.19 (p < 0.01). CONCLUSIONS: This modality was useful in identifying patients whose dyspnea was due to OAD. The ability to objectively and non-invasively measure these differences may prove clinically useful in distinguishing the operant physiology in patients presenting with acute dyspnea.


Assuntos
Auscultação/instrumentação , Diagnóstico por Imagem , Dispneia/diagnóstico , Pneumopatias Obstrutivas/diagnóstico , Adulto , Idoso , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Projetos Piloto , Espectrografia do Som , Vibração
4.
J Asthma ; 45(7): 575-8, 2008 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-18773329

RESUMO

BACKGROUND: Asthma is a disease of air flow obstruction. Transmitted sounds can be analyzed in detail and may shed light upon the physiology of asthma and how it changes over time. The goals of this study were to use a computerized analytic acoustic tool to evaluate respiratory sound patterns in asthmatic patients during acute attacks and after clinical improvement and to compare asthmatic profiles with those of normal individuals. METHODS: Respiratory sound analysis throughout the respiratory cycle was performed on 22 symptomatic asthma patients at the time of presentation to the emergency department (ED) and after clinical improvement. Fifteen healthy volunteers were analyzed as a control group. Vibrations patterns were plotted. Right and left lungs were analyzed separately. RESULTS: Asthmatic attacks were found to be correlated with asynchrony between lungs. In normal subjects, the inspiratory and expiratory vibration energy peaks (VEPs) occurred almost simultaneously in both lungs; the time interval between right and left expiratory VEPs was 0.006 +/- 0.012 seconds. In symptomatic asthmatic patients on admission, the time interval between right and left expiratory VEPs was 0.14 +/- 0.09 seconds and after clinical improvement the interval decreased to 0.04 +/- 0.04 seconds. Compared to healthy volunteers, asynchrony between two lungs was increased in asthmatics (p < 0.05). The asynchrony was significantly reduced after clinical improvement (p < 0.05). CONCLUSIONS: Respiratory sound analysis demonstrated significant asynchrony between right and left lungs in asthma exacerbations, a finding which, to our knowledge, has never been reported to date. The asynchrony is significantly reduced with clinical improvement following treatment.


Assuntos
Asma/fisiopatologia , Pulmão/fisiopatologia , Adulto , Estudos de Casos e Controles , Expiração/fisiologia , Feminino , Humanos , Inalação/fisiologia , Masculino , Pessoa de Meia-Idade , Sons Respiratórios , Índice de Gravidade de Doença
5.
Anesth Analg ; 107(4): 1243-7, 2008 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-18806034

RESUMO

BACKGROUND: Vibration response imaging (VRI) is a computer-based technology that creates a visual dynamic two-dimensional image of distribution of vibration within the lung during the respiratory process. The acoustic signals, recorded from 36 posteriorly positioned surface skin sensors, are transferred to a hardware board where several stages of filtering are applied to select a specific frequency band. The filtered output signal frequencies are presented as a gray-scale coded dynamic image, consisting of a series of 0.17 s frames, and as a table featuring the percentage contribution of each lung to the total vibration signal. METHODS: We describe the VRI technology in detail and examine images obtained from consecutive intensive care unit (ICU) patients with one diseased lung on chest radiograph. ICU patients with normal chest radiographs are presented as controls. Analysis of the image was performed by comparing the weighted pixel count analysis from both lungs. In this method, the pixels in the image were assigned values based on their grayscale color with the darker pixels assigned higher values. RESULTS: In patients with normal chest radiographs, the right and left lungs developed similarly in dynamic VRI images, and the percent lung vibrations from both sides were comparable (53%+/-12% and 47%+/-12%, respectively). In ICU patients with asymmetric lung disease, however, the percent lung vibrations from the diseased and nondiseased lungs were 27%+/-23% and 73%+/-23%, respectively (P<0.001). In patients with asymmetric lung disease (one lung has moderate to severe disease and the other appears normal or close to normal as per chest radiograph), the diseased lung usually appeared in VRI as irregular, smaller, and lighter in color (reduced vibration signal) when compared to the nonaffected lung. The weighted pixel count from diseased and nondiseased lungs were 33%+/-21% and 67%+/-21%, respectively (P < 0.003). CONCLUSION: The VRI technology may provide a radiation-free method for identifying and tracking of asymmetric lung parenchymal processes.


Assuntos
Pneumopatias/diagnóstico , Pulmão/fisiopatologia , Respiração , Processamento de Sinais Assistido por Computador , Vibração , Adulto , Feminino , Humanos , Unidades de Terapia Intensiva , Masculino , Pessoa de Meia-Idade , Radiografia Torácica , Respiração Artificial
6.
Crit Care ; 11(1): R26, 2007.
Artigo em Inglês | MEDLINE | ID: mdl-17316449

RESUMO

INTRODUCTION: There are several ventilator modes that are used for maintenance mechanical ventilation but no conclusive evidence that one mode of ventilation is better than another. Vibration response imaging is a novel bedside imaging technique that displays vibration energy of lung sounds generated during the respiratory cycle as a real-time structural and functional image of the respiration process. In this study, we objectively evaluated the differences in regional lung vibration during different modes of mechanical ventilation by means of this new technology. METHODS: Vibration response imaging was performed on 38 patients on assist volume control, assist pressure control, and pressure support modes of mechanical ventilation with constant tidal volumes. Images and vibration intensities of three lung regions at maximal inspiration were analyzed. RESULTS: There was a significant increase in overall geographical area (p < 0.001) and vibration intensity (p < 0.02) in pressure control and pressure support (greatest in pressure support), compared to volume control, when each patient served as his or her own control while targeting the same tidal volume in each mode. This increase in geographical area and vibration intensity occurred primarily in the lower lung regions. The relative percentage increases were 28.5% from volume control to pressure support and 18.8% from volume control to pressure control (p < 0.05). Concomitantly, the areas of the image in the middle lung regions decreased by 3.6% from volume control to pressure support and by 3.7% from volume control to pressure control (p < 0.05). In addition, analysis of regional vibration intensity showed a 35.5% relative percentage increase in the lower region with pressure support versus volume control (p < 0.05). CONCLUSION: Pressure support and (to a lesser extent) pressure control modes cause a shift of vibration toward lower lung regions compared to volume control when tidal volumes are held constant. Better patient synchronization with the ventilator, greater downward movement of the diaphragm, and decelerating flow waveform are potential physiologic explanations for the redistribution of vibration energy to lower lung regions in pressure-targeted modes of mechanical ventilation.


Assuntos
Pulmão/fisiologia , Ventilação Pulmonar , Respiração Artificial/métodos , Vibração , Acústica , Adulto , Idoso , Idoso de 80 Anos ou mais , Feminino , Humanos , Processamento de Imagem Assistida por Computador , Masculino , Pessoa de Meia-Idade , Pressão , Estatística como Assunto , Volume de Ventilação Pulmonar
7.
Protein Pept Lett ; 13(4): 357-61, 2006.
Artigo em Inglês | MEDLINE | ID: mdl-16712511

RESUMO

Purified human liver arylsulfatase A (ASA) as well as an ASA peptide (residues 28-39) were sulfated by tyrosyl protein sulfotransferase in vitro. The media, but not the cell lysate, of normal human fibroblasts contained a tyrosine sulfated protein (pI = 4.5-5.5). This protein was not present in either media or cell lysate of human fibroblasts lacking ASA protein. These results suggest that tyrosine sulfation facilitates secretion of ASA and that this may have pathophysiological consequences.


Assuntos
Cerebrosídeo Sulfatase/metabolismo , Tirosina/metabolismo , Meios de Cultura/química , Eletroforese em Gel de Poliacrilamida , Fibroblastos/metabolismo , Humanos , Focalização Isoelétrica , Fígado/enzimologia , Fragmentos de Peptídeos/metabolismo , Sulfotransferases/metabolismo , Radioisótopos de Enxofre
8.
Can J Anaesth ; 55(3): 172-6, 2008 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-18310628

RESUMO

PURPOSE: We describe the effect that inadvertent esophageal intubation has on the images and on the vibration distribution of vibration response imaging (VRI). CLINICAL FEATURES: Vibration response imaging (VRI) is a novel, non-invasive, computer-based technology that measures vibration energy of lung sounds during respiration and displays regional intensity, in both visual and graphic format. Vibration response images, obtained prior to tracheal intubation (spontaneous breathing) and during endotracheal ventilation using a controlled mode, resulted in evenly distributed vibrations throughout the patient's lungs. During inadvertent esophageal ventilation, however, the majority of vibrations were detected in the upper regions of the image, compared to those of the lower (60% vs 8%, respectively). During spontaneous breathing and endotracheal ventilation, the midclavicular column of sensors, located over the centre of each lung, detected more vibrations compared to either the medial or the axillary column of sensors. During inadvertent esophageal ventilation, more vibrations were detected by the medial column of sensors (nearest to the midline/esophagus); and fewer were detected by the sensors that were positioned more laterally. CONCLUSION: This report illustrates the potential for a visual image of distribution of lung vibration energy to differentiate endotracheal intubation from inadvertent esophageal intubation.


Assuntos
Diagnóstico por Computador/métodos , Esôfago , Intubação , Vibração , Idoso , Feminino , Humanos , Intubação Intratraqueal , Erros Médicos , Sons Respiratórios/fisiologia , Processamento de Sinais Assistido por Computador/instrumentação
9.
Alcohol Clin Exp Res ; 30(11): 1950-5, 2006 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-17067361

RESUMO

BACKGROUND: Arylsulfatase A (ASA) is an enzyme that catalyzes the degradation of sulfatides, a glycosphingolipid found in many tissues, but predominantly in myelin and kidney. Arylsulfatase A is 1 member of a family of sulfatases that is activated by a required co- or posttranslational modification with the oxidation of cysteine to formylglycine. This conversion requires a novel oxygenase mechanism that can be inhibited by reactive oxygen species. Ethanol is known to cause an increase in reactive oxygen species in the liver. Because of its effect on the levels of hepatic reactive oxygen species, we hypothesized that ethanol would cause a specific decrease of rat hepatic ASA activity levels. METHODS: Male Sprague-Dawley rats received ethanol-containing, Lieber-DeCarli liquid diets for 15 days, and control rats were pair-fed a liquid diet in which dextrose was isocalorically substituted for ethanol. RESULTS: Arylsulfatase A activity levels decreased in livers of animals receiving alcohol compared with control animals. No significant changes in ASA activity levels were observed in the cerebral cortex and kidney. Furthermore, ethanol did not have any significant effect on hexosaminidase activity in any of the tissues examined. CONCLUSION: Ethanol caused a tissue-specific decrease in hepatic ASA activity levels, but not hexosaminidase activity levels.


Assuntos
Depressores do Sistema Nervoso Central/farmacologia , Cerebrosídeo Sulfatase/metabolismo , Etanol/farmacologia , Fígado/enzimologia , Animais , Western Blotting , Depressores do Sistema Nervoso Central/sangue , Córtex Cerebral/efeitos dos fármacos , Córtex Cerebral/metabolismo , Densitometria , Eletroforese em Gel de Poliacrilamida , Etanol/sangue , Hexosaminidases/metabolismo , Processamento de Imagem Assistida por Computador , Fígado/efeitos dos fármacos , Masculino , Ratos , Ratos Sprague-Dawley , Espécies Reativas de Oxigênio/metabolismo
10.
PORTO ALEGRE; ARTMED - ARTES MEDICAS SUL; 3¦ ED; 2000. 606 p. ilus, tab.
Monografia em Pt | SES SP - Acervo Centro de Referência e Treinamento DST/AIDS | ID: crt-701

RESUMO

O GUIA PARA PROCEDIMENTOS DE ENFERMAGEM E UMA REFERENCIA RAPIDA PARA SUPORTE CLINICO DESTINADO AOS ESTUDANTES DE TODOS OS TIPOS DE PROGRAMAS EDUCACIONAIS E PRATICOS DE ENFERMAGEM NO AMBIENTE CLINICO. ESTE LIVRO EXPOE OS PASSOS -CHAVE NECESSARIOS PARA O EMPREGO DOS PRINCIPIOS DE ENFERMAGEM E FORNECE CONHECIMENTOS CRITERIOSOS PROPRIOS PARA O CUIDADO DO PACIENTE. SAO APRESENTADAS AQUI INFORMACOES SOBRE APROXIMADAMENTE 200 PROCEDIMENTOS. UM QUADRO DETALHADO DE CONTEUDOS E INDICE FOI SELECIONADO PARA UMA FACIL LOCALIZACAO DOS PROCEDIMENTOS. OS PROCEDIMENTOS CONTIDOS NOS 12 CAPITULOS DESTA OBRA ESTAO ORGANIZADOS NO FORMATO DO PROCESSO, COM PROCEDIMENTOS LISTADOS NO INICIO DE CADA CAPITULO PARA MAIOR CONVENIENCIA. A VISAO GERAL DO CAPITULO REVE CONCEITOS E PRINCIPIOS BASICOS. UMA LISTA DE POSSIVEIS DIAGNOSTICOS DE ENFERMAGEM ACOMPANHA CADA PROCEDIMENTO E ESTAO ORGANIZADOS DA SEGUINTE MANEIRA- 1 MATERIAL, 2 FINALIDADE(S) (INCLUINDO EXEMPLOS DE RESULTADO ESPERADO), 3 INVESTIGACAO, 4 DIAGNOSTICO DE ENFERMAGEM, 5 IDENTIFICACAO E PLANEJAMENTO DOS RESULTADOS (INCLUINDO PONTOS CENTRADOS NO PACIENTE, COM DESTAQUE ESPECIAL A CONSIDERACOES GENERALISTAS, PEDIATRICAS, GERIATRICAS, DO CUIDADO DOMICILIAR E ASPECTOS TRANSCULTURAIS), 6 IMPLEMENTACAO (ACAO COM JUSTIFICATIVAS), 7 AVALIACAO, 8 DOCUMENTACAO (INCLUINDO EXEMPLOS DE ANOTACOES).


Assuntos
Humanos , Masculino , Feminino , Adulto , Idoso , Criança , Adolescente , Pré-Escolar , Isolamento de Pacientes , Estresse Fisiológico , Organização e Administração , Preparações Farmacêuticas , 35147 , Higiene , Terapêutica , Descanso , Sono , Tração , Sulfato de Cálcio , Muletas , Cateterismo , Eliminação de Partículas Virais , Catéteres , Sonda de Prospecção , Terapêutica , Balanço Hidrológico , 52503 , Drenagem , Oxigenação , Palpação , Pressão Sanguínea , Sinais Vitais , Pesos e Medidas , Evolução Biológica , Empatia , Terapêutica , Comunicação , Guia , Enfermagem
SELEÇÃO DE REFERÊNCIAS
Detalhe da pesquisa