Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
Brachytherapy ; 22(4): 518-523, 2023.
Artigo em Inglês | MEDLINE | ID: mdl-37211487

RESUMO

PURPOSE: Coronary stents reduce IVBT radiation dose with a single layer by 10-30%. However, the impact of multiple stent layers and stent expansion remains unexplored. Individualized dose adjustments considering variations in stent layers and expansion could improve radiation delivery effectiveness. METHODS: EGSnrc was used to compute the delivered vessel wall dose in various IVBT scenarios. Stent effects were modeled for the stent density of 25%, 50%, and 75% with 1, 2, and 3 layers respectively. Doses were calculated at 1.75 to 5.00 mm away from the source center, normalized to 100% at 2 mm. RESULTS: Dose fall-off increased with increasing stent density. With a single layer, the dose at 2 mm from source fell from 100% of prescription to 92%, 83% and 73% at 25%, 50% and 75% density, respectively. The computed dose to points with increasing radial distance from the source decreased progressively with increasing stent layers. With three layers, at 75% stent density, the dose at 2 mm from source center fell to 38%. CONCLUSIONS: A schema for image-guided IVBT dose adjustment is described. While it would be an improvement over current standard of care, myriad factors remain to be addressed in a comprehensive effort to optimize IVBT.


Assuntos
Braquiterapia , Reestenose Coronária , Humanos , Braquiterapia/métodos , Radioisótopos , Stents
SELEÇÃO DE REFERÊNCIAS
Detalhe da pesquisa