Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
País como assunto
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
Ann Vasc Surg ; 56: 216-223, 2019 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-30500645

RESUMO

BACKGROUND: Aortic arch curvature can be straightened by endograft placement. However, different measurement methods with dissimilar follow-up and endografts have been published. The aim of this study was to corroborate, for the first time, the pliability of the Conformable Gore TAG Thoracic Endoprosthesis (W.L. Gore and Associates, Flagstaff, AZ, USA) into the aortic arch, using different image vector analysis. MATERIAL AND METHODS: We, retrospectively, analyzed patients primarily treated for thoracic aortic aneurysms and blunt traumatic aortic injuries by means of a Conformable Gore TAG Thoracic Endoprosthesis endograft proximally sealed into the aortic arch (zones Z1-Z3) in five different Spanish centers, between 2010 and 2017. The preoperative, one-month and six-month postoperative, computed tomography angiographies (CTAs) were obtained, creating accurate 3D center lumen line and external lumen line from the aortic valve to the renal arteries. Three different image analysis methods were used to compare modifications of the aortic curvature: first, segment analysis (angulations of the center lumen line when divided into seven precise segments, examining anterior-posterior, right-left, and cranial-caudal displacement), second, center lumen line analysis (bending of the center lumen line itself in seven definite points), and third, expected behavior (length of the endograft in the external lumen line). Two independent observers performed a blind analysis of all CTAs. Changes between preoperative and postoperative CTAs at one and six months are compared, and differences are viewed between cases sealed proximally (Z1-Z2) and distally (Z3) into the aortic arch. RESULTS: We analyzed 37 cases. At 1- and 6-month follow-ups, minimal changes occurred first in segment analysis (only a slight decrease of -2.0° in the XY plane at 10 cm from the brachiocephalic trunk at six-month follow-up was seen, P = 0.027). Second, center lumen line analysis again only showed negligible aortic curvature straightening (+3.5° at 10 cm from the brachiocephalic trunk at one month, P = 0.006, disappearing at six-month follow-up). Finally, good device length predictability was shown (interclass correlation coefficients: 0.995 and 0.994 at one and six months, P > 0.001). No differences were seen between cases proximally sealed into the proximal and distal aortic arch. CONCLUSIONS: Conformable Gore TAG Thoracic Endoprosthesis thoracic endograft showed a good pliability into the aortic arch and proximal thoracic aorta, with minimal changes in the aortic curvature after endograft placement in the short-term follow-up (up to six months). In addition, final endograft length into outer aortic curvature is highly predictable.


Assuntos
Aorta Torácica/diagnóstico por imagem , Aorta Torácica/cirurgia , Aneurisma da Aorta Torácica/diagnóstico por imagem , Aneurisma da Aorta Torácica/cirurgia , Aortografia/métodos , Implante de Prótese Vascular/instrumentação , Prótese Vascular , Angiografia por Tomografia Computadorizada/métodos , Procedimentos Endovasculares/instrumentação , Interpretação de Imagem Radiográfica Assistida por Computador/métodos , Lesões do Sistema Vascular/diagnóstico por imagem , Lesões do Sistema Vascular/cirurgia , Idoso , Idoso de 80 Anos ou mais , Aorta Torácica/lesões , Aorta Torácica/fisiopatologia , Aneurisma da Aorta Torácica/fisiopatologia , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Valor Preditivo dos Testes , Desenho de Prótese , Estudos Retrospectivos , Espanha , Fatores de Tempo , Resultado do Tratamento , Remodelação Vascular , Lesões do Sistema Vascular/fisiopatologia
SELEÇÃO DE REFERÊNCIAS
Detalhe da pesquisa