Biomed Instrum Technol ; 55(s3): 17-26, 2021 Mar 01.
Artigo em Inglês | MEDLINE | ID: mdl-34153999

RESUMO

Section 5.1.2 of ANSI/AAMI/ISO 11137-1 states that "the potential for induced radioactivity in product shall be assessed." This article describes how compliance with this requirement may be achieved using qualified test methods. Materials of consideration are conceptually discussed, and results of testing conducted on products processed with a 7.5-MeV X-ray irradiation process are provided. As X-ray becomes more widely used in healthcare sterilization, having standard assessment protocols for activation coupled with a shared database of material test results will benefit manufacturers seeking to utilize this innovative technology.


Assuntos
Radioatividade , Instalações de Saúde , Esterilização , Raios X