Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
Anal Bioanal Chem ; 410(20): 5085-5092, 2018 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-29909459

RESUMO

The treatment of some inborn metabolism errors requires cholesterol substitution therapy. Cholesterol plays a vital role in the human body. Therefore, the majority of cholesterol determination techniques are targeted to blood and blood serum. Nevertheless, cholesterol determination in food is important as well. In this paper, cholesterol determination using differential pulse voltammetry (DPV) in dairy products (e.g., milk, clotted cream, yogurt, butter, etc.) is reported with a novel nonenzymatic sensor based on diphosphonic acid of 1,4-diacetylglycoluril (DPADGU) as an electrode surface modifier. Stable anodic response was obtained from cholesterol on the modified carbon-based electrode. The sensor has high stability, sensitivity (20 µA mol L-1 cm-2), and a wide linear range from 1 up to 200 µM. The LOD and LOQ values are 1.5 and 5.1 µM, respectively. The developed methods were successfully applied to the above mentioned dairy products. Graphical abstract ᅟ.


Assuntos
Colesterol/análise , Técnicas Eletroquímicas/instrumentação , Análise de Alimentos/instrumentação , Técnicas Biossensoriais , Calibragem , Catálise , Colesterol/normas , Concentração de Íons de Hidrogênio , Padrões de Referência , Espectroscopia de Infravermelho com Transformada de Fourier
SELEÇÃO DE REFERÊNCIAS
Detalhe da pesquisa