1.
Endoscopy ; 50(12): 1205-1234, 2018 12.
Artigo em Inglês | MEDLINE | ID: mdl-30458567

RESUMO

This Position Statement from the European Society of Gastrointestinal Endoscopy (ESGE) and the European Society of Gastroenterology Nurses and Associates (ESGENA) sets standards for the reprocessing of flexible endoscopes and endoscopic devices used in gastroenterology. An expert working group of gastroenterologists, endoscopy nurses, chemists, microbiologists, and industry representatives provides updated recommendations on all aspects of reprocessing in order to maintain hygiene and infection control.


Assuntos
Desinfecção/métodos , Desinfecção/normas , Endoscópios/normas , Endoscopia Gastrointestinal/instrumentação , Contaminação de Equipamentos/prevenção & controle , Controle de Infecções/normas , Documentação/normas , Humanos , Saúde Ocupacional/normas , Esterilização/métodos , Esterilização/normas
2.
Endoscopy ; 49(11): 1098-1106, 2017 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-29036747

RESUMO

Patients should be informed about the benefits and risks of endoscopic retrograde cholangiopancreatography (ERCP)Only specially trained and competent personnel should carry out endoscope reprocessing.Manufacturers of duodenoscopes should provide detailed instructions on how to use and reprocess their equipment.In the case of modifications to their equipment, manufacturers should provide updated instructions for use.Detailed reprocessing protocols based on the manufacturer's instructions for use should clearly lay out the different reprocessing steps necessary for each endoscope model.Appropriate cleaning equipment should be used for duodenoscopes in compliance with the manufacturer's instructions for use. Only purpose-designed, endoscope type-specific, single-use cleaning brushes should be used, to ensure optimal cleaning. As soon as the endoscope is withdrawn from the patient, bedside cleaning should be performed, followed by leak testing, thorough manual cleaning steps, and automated reprocessing, in order to: · Remove debris from external and internal surfaces;. · Prevent any drying of body fluids, blood, or debris;. · Prevent any formation of biofilms.. In addition to the leak test, visual inspection of the distal end as well as regular maintenance of duodenoscopes should be performed according to the manufacturer's instructions for use, in order to detect any damage at an early stage.The entire reprocessing procedure in endoscope washer-disinfectors (EWDs) should be validated according to the European and International Standard, EN ISO 15883. Routine technical tests of EWDs should be performed according to the validation reports.Microbiological surveillance of a proportion of the department's endoscopes should be performed every 3 months, with the requirement that all endoscopes used in the unit are tested at least once a year.In the case of suspected endoscopy-related infection, the relevant device (e. g., endoscope, EWD) should be taken out of service until adequate corrective actions have been taken. Outbreaks should be managed by a multidisciplinary team, including endoscopy, hygiene, and microbiology experts, manufacturers, and regulatory bodies, according to national standards and/or laws. In the case of suspected multidrug-resistant organism (MDRO) outbreaks, close cooperation between the endoscopy unit and the clinical health provider is essential (including infection control departments and hospital hygienists).


Assuntos
Infecção Hospitalar/prevenção & controle , Descontaminação/métodos , Descontaminação/normas , Resistência a Múltiplos Medicamentos , Duodenoscópios/normas , Contaminação de Equipamentos/prevenção & controle , Infecção Hospitalar/microbiologia , Duodenoscópios/microbiologia , Humanos , Controle de Infecções/métodos
3.
Endoscopy ; 49(12): 1262-1275, 2017 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-29145674

RESUMO

1 Prerequisites. The clinical service provider should obtain confirmation from the endoscope washer-disinfector (EWD) manufacturer that all endoscopes intended to be used can be reprocessed in the EWD. 2 Installation qualification. This can be performed by different parties but national guidelines should define who has the responsibilities, taking into account legal requirements. 3 Operational qualification. This should include parametric tests to verify that the EWD is working according to its specifications. 4 Performance qualification. Testing of cleaning performance, microbiological testing of routinely used endoscopes, and the quality of the final rinse water should be considered in all local guidelines. The extent of these tests depends on local requirements. According to the results of type testing performed during EWD development, other parameters can be tested if local regulatory authorities accept this. Chemical residues on endoscope surfaces should be searched for, if acceptable test methods are available. 5 Routine inspections. National guidelines should consider both technical and performance criteria. Individual risk analyses performed in the validation and requalification processes are helpful for defining appropriate test frequencies for routine inspections.


Assuntos
Desinfecção/instrumentação , Desinfecção/normas , Endoscópios/microbiologia , Reutilização de Equipamento/normas , Controle de Qualidade , Desinfecção/métodos , Documentação , Endoscópios/normas , Contaminação de Equipamentos/prevenção & controle , Guias como Assunto , Estudos de Validação como Assunto
4.
GMS Hyg Infect Control ; 17: Doc14, 2022.
Artigo em Inglês | MEDLINE | ID: mdl-36157383

RESUMO

The SARS-CoV-2 pandemic illustrates the necessity of effective preventive measures for existing and newly emerging pathogens. When confronted with pathogens or spoilage agents, especially if they are not yet well studied, effective hygiene protocols are needed immediately. In the medical field, effective preventive measures are key to prevent vulnerable patients from infections. In production areas, effective hygiene measures are needed to protect goods from spoilage or microbial contamination. The European standardization framework established by the European Committee for Standardization (CEN) ensures that effective hygiene measures are available and can be immediately implemented when needed. Based on a broad portfolio of standards/laboratory tests, activity claims specifically addressing the special features of applications of antimicrobial formulations are substantiated. In this review, the concept of using standardized surrogate test organisms is explained, and the European standardized test approach to claim microbicidal and virucidal efficacy, the specificity of claims and their relevance for infection prevention measures is illustrated. Furthermore, relevance of the European Norm test methods is elucidated in the light of legal requirements. Finally, the review explains the systematics of the standardized methodological portfolio of CEN, Technical Committee 216, which is very useful when effective strategies for fighting or preventing microbial and viral induced infections, contaminations or spoilage are needed on an immediate basis.

SELEÇÃO DE REFERÊNCIAS