RESUMO

A new type of electromechanical total artificial heart (TAH) based on circular rolling-cylinder mechanism was developed to overcome critical problems in motor-driven artificial hearts such as large size and difficulties in fitting the heart to atrial remnants and arterial vessels. Its performance and reliability were evaluated in mock circulation and in an animal implant experiment. The total weight and volume of the pump is 650 g and 600 mL, respectively. This new pump was implanted in a calf for total heart replacement and 96 h of survival was achieved. The whole system, including pump, controller, and control algorithm performed well enough to improve the prospect of eventual clinical application of our TAH system.


Assuntos
Coração Artificial , Animais , Circulação Sanguínea/fisiologia , Feminino , Modelos Cardiovasculares , Desenho de Prótese , Ovinos
2.
ASAIO Trans ; 36(3): M249-51, 1990.
Artigo em Inglês | MEDLINE | ID: mdl-2252669

RESUMO

An electromechanical moving-actuator TAH was developed and implanted for 5 days in a calf experiment. The efficiency and durability were further improved using pendulous motion of the actuator. This new pendulum pump has a circular motion produced by gear mechanisms connected to the motor. It has a height (from valves to housing bottom) of 7 cm and centrally directed inlets and outlets for good fittability. Active filling (-10 mmHg) inside inner blood sacs was produced by the suction effects of the outer sacs attached to a moving actuator. Two acute animal experiments were performed using the pendulum pump with a Coridale sheep (50 kg weight), and a male calf (70 kg weight). A test was made of 30 minutes of tether-free operation of the implanted pendulum pump inside the small chest of the calf. The pump shows the engineering feasibility of the implantable electrical TAH inside a human chest cavity.


Assuntos
Coração Artificial , Hemodinâmica/fisiologia , Animais , Pressão Sanguínea/fisiologia , Bovinos , Desenho de Equipamento , Masculino , Modelos Cardiovasculares