RESUMO
OBJECTIVES: The purpose of this study was to evaluate the safety and efficacy of valve-in-valve (ViV) transcatheter aortic valve replacement (TAVR) for stentless bioprosthetic aortic valves (SBAVs) and to identify predictors of adverse events. BACKGROUND: ViV TAVR in SBAVs is associated with unique technical challenges and risks. METHODS: Clinical records and computer tomographic scans were retrospectively reviewed for procedural complications, predictors of coronary obstruction, mortality, and echocardiographic results. RESULTS: Among 66 SBAV patients undergoing ViV TAVR, mortality was 2 of 66 patients (3.0%) at 30 days and 5 of 52 patients (9.6%) at 1 year. At 1 year, left ventricular end-systolic dimension was decreased versus baseline (median [interquartile range (IQR)]: 3.0 [2.6 to 3.6] cm vs. 3.7 [3.2 to 4.4] cm; p < 0.001). Coronary occlusion in 6 of 66 procedures (9.1%) resulted in myocardial infarction in 2 of 66 procedures (3.0%). Predictors of coronary occlusion included subcoronary implant technique compared with full root replacement (6 of 31, 19.4% vs. 0 of 28, 0%; p = 0.01), short simulated radial valve-to-coronary distance (median [IQR]: 3.4 [0.0 to 4.6] mm vs. 4.6 [3.2 to 6.2] mm; p = 0.016), and low coronary height (7.8 [5.8 to 10.0] mm vs. 11.6 [8.7 to 13.9] mm; p = 0.003). Coronary arteries originated <10 mm above the valve leaflets in 34 of 97 unobstructed coronary arteries (35.1%). CONCLUSIONS: TAVR in SBAVs is frequently associated with high-risk coronary anatomy but can be performed with a low risk of death and myocardial infarction, resulting in favorable ventricular remodeling. A subcoronary surgical approach is associated with an increased risk of coronary obstruction.