RESUMO
This study compared the performance of 3 automated immunoassays, Architect® (Abbott), Immulite® (Siemens) and Liaison® (Diasorin), for Epstein-Barr virus (EBV) serology. Ninety-one serum samples collected in Amiens University Hospital were analyzed for the presence of Viral Capsid Antigen (VCA) IgG and IgM and Epstein-Barr Nuclear Antigen (EBNA) IgG. The agreement between the 3 assays was calculated for each marker individually and for determination of the EBV profile, based on interpretation of the combination of these 3 EBV markers. Although similar results were obtained with Architect® and Liaison®, several discordant results were observed with Immulite®, particularly for EBNA IgG. A large number of EBNA IgG-positive results were observed, which interfered with interpretation of the EBV profile. In contrast, Immulite® performed similarly to the 2 other assays for detection of VCA IgM.