Your browser doesn't support javascript.
loading
Enantioselective analysis of methadone in saliva by liquid chromatography-mass spectrometry.
Ortelli, D; Rudaz, S; Chevalley, A F; Mino, A; Deglon, J J; Balant, L; Veuthey, J L.
  • Ortelli D; Laboratory of Pharmaceutical Analytical Chemistry, University of Geneva, Switzerland.
J Chromatogr A ; 871(1-2): 163-72, 2000 Feb 25.
Article en En | MEDLINE | ID: mdl-10735296
Saliva was tested and evaluated as a biological matrix for methadone (Mtd) monitoring. Conventional method using a narrow bore C18 column, and an enantioselective method using a narrow bore alpha1-acid glycoprotein column, were developed using liquid chromatography coupled with a mass spectromeric (MS) detector. After optimisation of MS conditions by flow injection analysis, selected ion monitoring detection was used to enhance sensitivity. The total Mtd concentration and the enantiomeric ratio in saliva were validated using an experimental design. The methods were applied to samples provided by heroin addicts undergoing a Mtd treatment. Results on total Mtd determination showed a very poor correlation between saliva and serum, whereas the enantiomeric ratios of Mtd gave a very good one.
Asunto(s)
Search on Google
Banco de datos: MEDLINE Asunto principal: Saliva / Cromatografía Liquida / Metadona Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: En Año: 2000 Tipo del documento: Article
Search on Google
Banco de datos: MEDLINE Asunto principal: Saliva / Cromatografía Liquida / Metadona Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: En Año: 2000 Tipo del documento: Article