Your browser doesn't support javascript.
loading
Dynamic model of amperometric biosensors. Characterisation of glucose biosensor output.
Rinken, T; Tenno, T.
  • Rinken T; Institute of Physical Chemistry, Tartu University, Estonia. toonika@ut.ee
Biosens Bioelectron ; 16(1-2): 53-9, 2001 Jan.
Article en En | MEDLINE | ID: mdl-11261853
ABSTRACT
An integrated model for the characterisation of the output signal course of oxidase-bound amperometric biosensors is presented and evaluated in the case of glucose biosensors. This model integrates two earlier proposed models, the model of oxygen transducer-based biosensors, allowing the prediction of steady state parameters from the transient response and the dynamic signal lag model, characterising the electrochemical diffusion-limited sensors. The integrated model allows the characterisation of the whole biosensor signal output, originating from the output curve itself with errors less than 3% and no need to determine the system's geometrical, diffusion and partition parameters.
Asunto(s)
Search on Google
Banco de datos: MEDLINE Asunto principal: Técnicas Biosensibles / Glucosa Tipo de estudio: Prognostic_studies Idioma: En Año: 2001 Tipo del documento: Article
Search on Google