Your browser doesn't support javascript.
loading
CR-39 acceptance test and optimisation for fast neutron dosimetry applications.
Fantuzzi, E; Morelli, B; Falangi, G; Patrizii, L; Togo, V.
  • Fantuzzi E; ENEA, Institute for Radiation Protection, Bologna, Italy. fantuzzi@bologna.enea.it
Radiat Prot Dosimetry ; 101(1-4): 573-8, 2002.
Article en En | MEDLINE | ID: mdl-12382815
ABSTRACT
The ENEA fast neutron dosemeter is based on a planar PADC (poly allyl diglycol carbonate) placed in a polyethylene holder. The CR-39 (registered trademark of PPG Industries Inc.) material, produced by Intercast Europe S.p.A., has been used in the routine of the Individual Monitoring Service (IMS) since 1998. Since then, acceptance tests on average sheet background track density and sheet neutron sensitivity have been made on new batches as a quality control within a quality assurance programme of the IMS of ENEA-Institute for Radiation Protection (IRP). Dosemeters were irradiated with a 241Am-Be source at ENEA-IRP and processed through a chemical etching procedure (pre-etching with 40% KOH water solution 6.25 N and 60% ethyl alcohol at 70 degrees C followed by 12 h of etching in 6.25 N KOH water solution). In this paper we present the analysis of acceptance testing data for more than 30 sheets of CR-39 plastic produced in 1998, 1999 and 2000. Moreover, we compare the performance of sheets of CR-39 of standard composition with that of sheets of CR-39 with the addition of DOP (dioctylphthalate), in different concentrations, on the hasis of average background density, neutron sensitivity and background fluctuation that limit the lower detectable dose. This study demonstrates the need for acceptance tests to assure the quality of the dosimetric performance of these dosemeters, which is considerably dependent on the quality of the CR-39 plastic.
Asunto(s)
Search on Google
Banco de datos: MEDLINE Asunto principal: Monitoreo de Radiación / Neutrones Rápidos Idioma: En Año: 2002 Tipo del documento: Article
Search on Google
Banco de datos: MEDLINE Asunto principal: Monitoreo de Radiación / Neutrones Rápidos Idioma: En Año: 2002 Tipo del documento: Article