Analytical performance and method comparison study of the total homocysteine fluorescence polarization immunoassay (FPIA) on the AxSYM analyzer.
Lonati, Silvia; Novembrino, Cristina; Ippolito, Silvia; Accinni, Roberto; Galli, Claudio; Troonen, Hugo; Campolo, Jonica; Della Noce, Cinzia; Lunghi, Giovanna; Catena, Fabrizia Bamonti.
  • Lonati S; Dipartimento Scienze Mediche, Università degli Studi, IRCCS Ospedale Maggiore Milano, Milan, Italy.
Clin Chem Lab Med ; 42(2): 228-34, 2004 Feb.
Article en En | MEDLINE | ID: mdl-15061366
ABSTRACT
A fluorescence polarization immunoassay (FPIA) has been commercially released for routine large-scale testing of total homocysteine (tHcy) on the AxSYM analyzer. We evaluated the analytical performance of the AxSYM tHcy FPIA and compared it with the well established high-performance liquid chromatography (HPLC) and IMx tHcy FPIA methods. Homocysteine concentrations were measured by AxSYM and IMx tHcy FPIA and by a rapid isocratic HPLC method with fluorescence detection. Coefficient of variation (CV) of total imprecision for AxSYM tHcy was < or = 5%, mean dilution recovery 102%, analytical sensitivity 0.70 micromol/l and linearity was good up to 18 dilution. Spearman rank correlations, rho, were 0.83 (p < 0.0001) for AxSYM vs. HPLC, 0.97 (p < 0.0001) for AxSYM vs. IMx and 0.83 (p < 0.0001) for IMx vs. HPLC. Passing and Bablok regression Y-intercepts and slopes were 2.944/0.937 (AxSYM vs. HPLC), -0.367/ 1.142 (AxSYM vs. IMx) and 2.632/0.805 (IMx vs. HPLC). Corresponding mean differences (AxSYM-Comparison Assay) recorded over a 5-50 micromol/l measured range were 1.80, -0.73 and 2.53 micromol/l. AxSYM tHcy FPIA's first rate precision, supported by the complete automation of the AxSYM analyzer, makes it fit for routine use and suitable for laboratories requiring homocysteine high-throughput testing capabilities.
Asunto(s)
Search on Google
Banco de datos: MEDLINE Asunto principal: Análisis Químico de la Sangre / Inmunoensayo de Polarización Fluorescente / Homocisteína Tipo de estudio: Evaluation_studies Límite: Humans Idioma: En Año: 2004 Tipo del documento: Article
Search on Google
Banco de datos: MEDLINE Asunto principal: Análisis Químico de la Sangre / Inmunoensayo de Polarización Fluorescente / Homocisteína Tipo de estudio: Evaluation_studies Límite: Humans Idioma: En Año: 2004 Tipo del documento: Article