Your browser doesn't support javascript.
loading
Measurement of the source size of a 6- and 18-MV radiotherapy linac.
Loewenthal, E; Loewinger, E; Bar-Avraham, E; Barnea, G.
  • Loewenthal E; Clinical Oncology and Radiation Therapy, Hadasa Hospital, Ein Kerem, Israel.
Med Phys ; 19(3): 687-90, 1992.
Article en En | MEDLINE | ID: mdl-1508108
ABSTRACT
The quality of portal imaging is strongly affected by the source size of the radiotherapy machine. The effective source size of the dual-energy clinac 1800 (6 and 18 MV) was measured with a 50 microns wide and 120 mm long slit formed by two tungsten-copper alloy blocks. A series of slit images were obtained by translating the slit horizontally. The images were analyzed using a microdensitometer. The measured data was simulated using an analytical model of the source and its size was derived by a best-fit analysis. For both energies the FWHM was found to be 1.5 +/- 0.1 mm.
Asunto(s)
Search on Google
Banco de datos: MEDLINE Asunto principal: Radioterapia Límite: Humans Idioma: En Año: 1992 Tipo del documento: Article
Search on Google
Banco de datos: MEDLINE Asunto principal: Radioterapia Límite: Humans Idioma: En Año: 1992 Tipo del documento: Article