Your browser doesn't support javascript.
loading
The reference ear modeling method for internally feedback controlled digital hearing aid chip.
Kim, Sunyoung; Lee, Seung Jin; Cho, Namjun; Song, Seong-Jun; Yoo, Hoi-Jun.
  • Kim S; Korea Advanced Institute of Science and technology, 373-1, Guseong-dong, Yuseong-gu, Daejeon, 305-701, Republic of Korea. sunyoung@eeinfo.kaist.ac.kr
Article en En | MEDLINE | ID: mdl-18003331
ABSTRACT
A reference ear modeling method for the real-time feedback controlled digital hearing aid chip is proposed and implemented. In order to reduce the modeling complexity and enhance the programmability, new ear modeling method using the acoustic filter theory is adopted to the digital hearing aid. To achieve the fully internal gain fitting and verification system, the responses from the damaged ear and the reference ear model are compared and the new gain parameters are processed for the multi-channel DSP. The digital hearing aid chip with reference ear model is fabricated in 0.18 microm CMOS technology, has a core area of 3.1 mm x 1.2 mm and dissipates less than 120 muA.
Asunto(s)
Search on Google
Banco de datos: MEDLINE Asunto principal: Espectrografía del Sonido / Biorretroalimentación Psicológica / Procesamiento de Señales Asistido por Computador / Oído / Audición / Audífonos / Modelos Biológicos Tipo de estudio: Evaluation_studies Idioma: En Año: 2007 Tipo del documento: Article
Search on Google
Banco de datos: MEDLINE Asunto principal: Espectrografía del Sonido / Biorretroalimentación Psicológica / Procesamiento de Señales Asistido por Computador / Oído / Audición / Audífonos / Modelos Biológicos Tipo de estudio: Evaluation_studies Idioma: En Año: 2007 Tipo del documento: Article