A stable isotope breath test with a standard meal for abnormal gastric emptying of solids in the clinic and in research.
Szarka, Lawrence A; Camilleri, Michael; Vella, Adrian; Burton, Duane; Baxter, Kari; Simonson, Julie; Zinsmeister, Alan R.
  • Szarka LA; Clinical Enteric Neuroscience Translational and Epidemiological Research, College of Medicine, Mayo Clinic, Rochester, Minnesota 55905, USA. szarka.lawrence@mayo.edu
Clin Gastroenterol Hepatol ; 6(6): 635-643.e1, 2008 Jun.
Article en En | MEDLINE | ID: mdl-18406670
ABSTRACT
BACKGROUND &

AIMS:

The aim of this study was to validate a [13C]-Spirulina platensis gastric emptying (GE) breath test (GEBT) with a standardized meal.

METHODS:

Thirty-eight healthy volunteers and 129 patients with clinically suspected delayed GE underwent measurements at 45, 90, 120, 150, 180, and 240 minutes after a 238 kcal meal labeled test with 100 mg [13C]-S platensis and 0.5 mCi 99mTc. We established normal ranges for scintigraphy with this test meal, intraindividual and interindividual coefficients of variation (COVs), and the ability of the [13C] GEBT breath percent dose excreted *1000 values to predict scintigraphic half-life and to categorize GE as delayed, normal, or accelerated.

RESULTS:

In health, the 10th and 90th percentiles of half-life for scintigraphic GE with this meal were 52 and 86 minutes; intraindividual COVs for scintigraphy and the GEBT were, respectively, 31% and 27% at 45 minutes, 17% and 21% at 90 minutes, 13% and 16% at 120 minutes, 10% and 13% at 150 minutes, and 8% and 12% at 180 minutes. Interindividual COVs at each time for the [13C] GEBT and scintigraphy were typically approximately 1%-4% lower than intraindividual COVs. Individual breath samples at 45, 150, and 180 minutes predicted GE category; at 80% specificity, 45- and 180-minute samples combined were 93% sensitive to identify accelerated GE, and 150- and 180-minute combined were 89% sensitive for delayed GE.

CONCLUSIONS:

[13C]-S platensis GEBT is as reproducible as scintigraphy; imprecision with both tests reflects physiologic variation. With 4 breath samples, this method with an off-the-shelf meal is valid to assess GE in clinic and in research.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Gastropatías / Pruebas Respiratorias / Isótopos de Carbono / Vaciamiento Gástrico Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Año: 2008 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Gastropatías / Pruebas Respiratorias / Isótopos de Carbono / Vaciamiento Gástrico Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Año: 2008 Tipo del documento: Article