Your browser doesn't support javascript.
loading
Guided wave phase velocity measurement using multi-emitter and multi-receiver arrays in the axial transmission configuration.
Minonzio, Jean-Gabriel; Talmant, Marilyne; Laugier, Pascal.
  • Minonzio JG; UPMC Univ Paris 06, UMR 7623, LIP, 15 Rue de l'École de Medicine, F-75005 Paris, France. jean-gabriel.minonzio@upmc.fr
J Acoust Soc Am ; 127(5): 2913-9, 2010 May.
Article en En | MEDLINE | ID: mdl-21117742
ABSTRACT
This paper is devoted to a method of extraction of guided waves phase velocities from experimental signals. Measurements are performed using an axial transmission device consisting of a linear arrangement of emitters and receivers placed on the surface of the inspected specimen. The technique takes benefit of using both multiple emitters and receivers and is validated on a reference wave guide. The guided mode phase velocities are obtained using a projection in the singular vectors basis. The singular vectors are determined by the singular values decomposition (SVD) of the response matrix between the two arrays in the frequency domain. This technique enables to recover accurately guided wave phase velocity dispersion curves. The SVD based approach was designed to overcome limitations of spatio-temporal Fourier transform for receiver array of limited spatial extent as in the case of clinical assessment of cortical bone in axial transmission.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Transductores / Ultrasonido / Huesos / Procesamiento de Señales Asistido por Computador / Modelos Teóricos Tipo de estudio: Diagnostic_studies / Prognostic_studies Idioma: En Año: 2010 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Transductores / Ultrasonido / Huesos / Procesamiento de Señales Asistido por Computador / Modelos Teóricos Tipo de estudio: Diagnostic_studies / Prognostic_studies Idioma: En Año: 2010 Tipo del documento: Article