ABSTRACT
Active electronic dosemeters using counting techniques are used for radioprotection purposes in pulsed radiation fields in X-ray diagnostics or therapy. The disadvantage of the limited maximum measurable dose rate becomes significant in these radiation fields and leads to some negative effects. In this study, a set of relevant parameters for a dosemeter is described, which can be used to decide whether it is applicable in a given radiation field or not. The determination of these relevant parameters-maximum measurable dose rate in the radiation pulse, dead time of the dosemeter, indication per counting event and measurement cycle time-is specified. The results of the first measurements on the determination of these parameters for an electronic personal dosemeter of the type Thermo Fisher Scientific EPD Mk2 are shown.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Radiación Ionizante / Protección Radiológica / Monitoreo de Radiación / Exposición Profesional Límite: Humans Idioma: En Año: 2012 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Radiación Ionizante / Protección Radiológica / Monitoreo de Radiación / Exposición Profesional Límite: Humans Idioma: En Año: 2012 Tipo del documento: Article