Breast dynamic contrast-enhanced examinations with fat suppression: are contrast-agent uptake curves affected by magnetic field inhomogeneity?
Schmidt, Maria A; Borri, M; Scurr, E; Ertas, G; Payne, G; O'Flynn, E; Desouza, N; Leach, M O.
  • Schmidt MA; Royal Marsden NHS Foundation Trust and Institute of Cancer Research, MRI Unit, Downs Rd, Sutton SM2 5 PT, UK. maria.schmidt@icr.ac.uk
Eur Radiol ; 23(6): 1537-45, 2013 Jun.
Article en En | MEDLINE | ID: mdl-23242002
ABSTRACT

OBJECTIVES:

To investigate the effect of magnetic field heterogeneity in breast dynamic contrast-enhanced examinations with fat saturation (DCE-FS).

METHODS:

The magnetic field was mapped over the breasts in ten patients. DCE-FS was undertaken at 1.5 T with fast spoiled gradient echoes and spectrally selective fat saturation. Signal intensity was calculated for T1 values 25-1,200 ms both on and off resonance, and results were verified with a test object. Clinical examinations were evaluated for the predicted effects of field heterogeneity.

RESULTS:

Magnetic field was found to vary by 3.6 ± 1.2 ppm over the central transaxial slice and 5.1 ± 1.5 over the whole breast volume (mean ± standard deviation). Computer simulations predict a reduction in the dynamic range if field heterogeneity leads to unintended water suppression, and distortion to CA uptake curves due to fat suppression failure (for fat containing pixels). A compromise between dynamic range and fat saturation performance is required. Both water suppression and fat suppression failure are apparent in clinical examinations.

CONCLUSION:

Magnetic field heterogeneity is likely to reduce the sensitivity of DCE-FS by distorting the CA uptake curves because of fat suppression failure (for fat containing pixels) and by reducing the dynamic range because of unintended water suppression. KEY POINTS • Magnetic field heterogeneity is significant in breast magnetic resonance. • Contrast-agent uptake curves are distorted by a non-uniform magnetic field. • Radiologist must be aware of possibility of distortion to interpret uptake curves correctly. • Compromise between fat suppression and dynamic range is required.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Mama / Neoplasias de la Mama / Medios de Contraste Tipo de estudio: Prognostic_studies Límite: Female / Humans Idioma: En Año: 2013 Tipo del documento: Article