[Identification of Streptococcus mutans in carious patients' saliva samples by MALDI-TOF mass spectrometry].
Ren, Wen; Chen, Feng; Zhang, Yi-fei; Zhang, Qian; Wang, Xiao-yan; Liu, Ying-yi; Yuan, Chong-yang; Ma, Qing-wei; Xu, Tao; Zheng, Shu-guo.
  • Ren W; Department of Preventive Dentistry, Peking University School and Hospital of Stomatology, Beijing 100081, China.
  • Chen F; Central Laboratory, Peking University School and Hospital of Stomatology, Beijing 100081, China.
  • Zhang YF; Central Laboratory, Peking University School and Hospital of Stomatology, Beijing 100081, China.
  • Zhang Q; Central Laboratory, Peking University School and Hospital of Stomatology, Beijing 100081, China.
  • Wang XY; Department of Endodontics, Peking University School and Hospital of Stomatology, Beijing 100081, China.
  • Liu YY; Department of Endodontics, Peking University School and Hospital of Stomatology, Beijing 100081, China.
  • Yuan CY; Department of Endodontics, Peking University School and Hospital of Stomatology, Beijing 100081, China.
  • Ma QW; Academy for Advanced Interdisciplinary Studies, Peking University, Beijing 100871, China.
  • Xu T; Department of Preventive Dentistry, Peking University School and Hospital of Stomatology, Beijing 100081, China.
  • Zheng SG; Department of Preventive Dentistry, Peking University School and Hospital of Stomatology, Beijing 100081, China.
Beijing Da Xue Xue Bao Yi Xue Ban ; 46(1): 25-9, 2014 Feb 18.
Article en Zh | MEDLINE | ID: mdl-24535342
ABSTRACT

OBJECTIVE:

To identify Streptococcus mutans (S. mutans) in carious patients' saliva using matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS), and to establish a faster and more accurate method to identify S. mutans.

METHODS:

In this study, a total of 90 carious patients from Department of Endodontics of Peking University School of Stomatology were recruited. All these patients' saliva was collected. After extracting the protein of the samples, they were identified by MALDI-TOF MS. The composite profile was analyzed using BioExplorer 1.0 software. The scores ≥ 25 were considered as S. mutans, whereas the scores <25 were as considered as non S. mutans. Finally, these results were compared with 16S rDNA sequencing to figure out the sensitivity and concordance rate, respectively.

RESULTS:

The sensitivity of MALDI-TOF MS was 96.0%, and the concordance rate compared with 16S rDNA sequencing was as high as 98.7%.

CONCLUSION:

MALDI-TOF MS is high throughput, rapid and easy to perform in comparison to other conventional methods. It has a high sensivity and concordance rate. Thus, MALDI-TOF MS can serve as an effective tool for identification of S. mutans.