Portable lactate analyzer for measuring lactate in cerebrospinal fluid (CSF) and plasma - method-comparison evaluations.
Almeida, Sérgio Monteiro de; Marquie-Beck, Jennifer; Bhatt, Archana; Letendre, Scott; McCutchan, Allen; Ellis, Ron.
  • Almeida SM; Departments of Neurosciences, HIV Neurobehavioral Research Center, University of California, San Diego, USA.
  • Marquie-Beck J; Departments of Neurosciences, HIV Neurobehavioral Research Center, University of California, San Diego, USA.
  • Bhatt A; Departments of Psychiatry, HIV Neurobehavioral Research Center, University of California, San Diego, USA.
  • Letendre S; School of Medicine, HIV Neurobehavioral Research Center, University of California, San Diego, USA.
  • McCutchan A; School of Medicine, HIV Neurobehavioral Research Center, University of California, San Diego, USA.
  • Ellis R; Departments of Neurosciences, HIV Neurobehavioral Research Center, University of California, San Diego, USA.
Arq Neuropsiquiatr ; 72(7): 500-5, 2014 Jul.
Article en En | MEDLINE | ID: mdl-25054981
ABSTRACT
UNLABELLED Increased plasma lactate levels can indicate the presence of metabolic disorders in HIV infected individuals.

OBJECTIVE:

To determine whether a portable analyzer is valid for measuring cerebrospinal fluid (CSF) and plasma lactate levels in HIV infected individuals.

METHOD:

CSF and plasma were collected from 178 subjects. Samples tested by the Accutrend portable analyzer were compared to those tested by a reference device (SYNCHRON LX 20).

RESULTS:

The portable analyzer had in plasma sensitivity of 0.95 and specificity 0.87. For CSF the specificity was 0.95; the sensitivity 0.33; the negative predictive value was 95% and the positive predictive value 33%.

CONCLUSIONS:

These findings support the validity of the portable analyzer in measuring lactate concentrations in CSF that fall within the normal range. The relatively poor positive predictive value indicates that a result above the reference range may represent a "false positive test", and should be confirmed by the reference device before concluding abnormality.