The Value of Quality Improvement Process in the Detection and Correction of Common Errors in Echocardiographic Hemodynamic Parameters in a Busy Echocardiography Laboratory.
Fanari, Zaher; Choudhry, Usman I; Reddy, Vivek K; Eze-Nliam, Chete; Hammami, Sumaya; Kolm, Paul; Weintraub, William S; Marshall, Erik S.
  • Fanari Z; Division of Cardiology, Christiana Care Health System, Newark, Delaware.
  • Choudhry UI; Division of Cardiology, Christiana Care Health System, Newark, Delaware.
  • Reddy VK; Division of Cardiology, Christiana Care Health System, Newark, Delaware.
  • Eze-Nliam C; Division of Cardiology, Christiana Care Health System, Newark, Delaware.
  • Hammami S; Division of Cardiology, Christiana Care Health System, Newark, Delaware.
  • Kolm P; Value Institute, Christiana Care Health System, Newark, Delaware.
  • Weintraub WS; Division of Cardiology, Christiana Care Health System, Newark, Delaware.
  • Marshall ES; Value Institute, Christiana Care Health System, Newark, Delaware.
Echocardiography ; 32(12): 1778-89, 2015 Dec.
Article en En | MEDLINE | ID: mdl-26033297
BACKGROUND: Accurate assessment of cardiac structures, ventricular function, and hemodynamics is essential for any echocardiographic laboratory. Quality improvement (QI) processes described by the American Society of Echocardiography (ASE) and the Intersocietal Commission (IAC) should be instrumental in reaching this goal. METHODS: All patients undergoing transthoracic echocardiogram (TTE) followed by cardiac catheterization within 24 hours at Christiana Care Health System in 2011 and 2012 were identified, with 126 and 133 cases, respectively. Hemodynamic parameters of diastolic function and pulmonary artery systolic pressure (PASP) on TTE correlated poorly with catheterization in 2011. An educational process was developed and implemented at quarterly QI meetings based on ASE and IAC recommendations to target frequently encountered errors and provide methods for improved performance. The hemodynamic parameters were then reexamined in 2012 postintervention. RESULTS: Following the QI process, there was significant improvement in the correlation between invasive and echocardiographic hemodynamic measurements in both systolic and diastolic function, and PASP. This reflected in significant better correlations between echo and cath LVEF [R = 0.88, ICC = 0.87 vs. R = 0.85, ICC = 0.85; P < 0.001], average E/E' and of left ventricle end-diastolic pressure (LVEDP) [R = 0.62 vs. R = 0.09, P = 0.006] and a better correlation for PASP [R = 0.77, ICC = 0.77 vs. R = 0.30, ICC = 0.31; P = 0.05] in 2012 compared to 2011. CONCLUSION: The QI process, as recommended by ASE and IAC, can allow for identification as well as rectification of quality issues in a large regional academic medical center hospital.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Laboratorios de Hospital / Ecocardiografía / Aumento de la Imagen / Guías de Práctica Clínica como Asunto / Errores Diagnósticos / Mejoramiento de la Calidad Tipo de estudio: Diagnostic_studies / Evaluation_studies / Guideline / Prognostic_studies Límite: Humans País como asunto: America do norte Idioma: En Año: 2015 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Laboratorios de Hospital / Ecocardiografía / Aumento de la Imagen / Guías de Práctica Clínica como Asunto / Errores Diagnósticos / Mejoramiento de la Calidad Tipo de estudio: Diagnostic_studies / Evaluation_studies / Guideline / Prognostic_studies Límite: Humans País como asunto: America do norte Idioma: En Año: 2015 Tipo del documento: Article