Your browser doesn't support javascript.
loading
Planning for mARC treatments with the Eclipse treatment planning system.
Sarkar, Vikren; Huang, Long; Rassiah-Szegedi, Prema; Zhao, Hui; Huang, Jessica; Szegedi, Martin; Salter, Bill J.
  • Sarkar V; University of Utah. vikren.sarkar@gmail.com.
J Appl Clin Med Phys ; 16(2): 5351, 2015 Mar 08.
Article en En | MEDLINE | ID: mdl-26103202
ABSTRACT
While modulated arc (mARC) capabilities have been available on Siemens linear accelerators for almost two years now, there was, until recently, only one treatment planning system capable of planning these treatments. The Eclipse treatment planning system now offers a module that can plan for mARC treatments. The purpose of this work was to test the module to determine whether it is capable of creating clinically acceptable plans. A total of 23 plans were created for various clinical sites and all plans delivered without anomaly. The average 3%/3 mm gamma pass rate for the plans was 98.0%, with a standard deviation of 1.7%. For a total of 14 plans, an equivalent static gantry IMRT plan was also created to compare delivery time. In all but two cases, the mARC plans delivered significantly faster than the static gantry plan. We have confirmed the successful creation of mARC plans that are deliverable with high fidelity on an ARTISTE linear accelerator, thus demonstrating the successful implementation of the Eclipse mARC module.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Planificación de la Radioterapia Asistida por Computador / Radiocirugia / Radioterapia de Intensidad Modulada / Neoplasias Tipo de estudio: Etiology_studies Límite: Humans Idioma: En Año: 2015 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Planificación de la Radioterapia Asistida por Computador / Radiocirugia / Radioterapia de Intensidad Modulada / Neoplasias Tipo de estudio: Etiology_studies Límite: Humans Idioma: En Año: 2015 Tipo del documento: Article