Experimental validation of a flat punch indentation methodology calibrated against unconfined compression tests for determination of soft tissue biomechanics.
Delaine-Smith, R M; Burney, S; Balkwill, F R; Knight, M M.
  • Delaine-Smith RM; School of Engineering and Materials Science, Institute of Bioengineering, Queen Mary University of London, Mile End, London E1 4NS, UK; Centre for Cancer and Inflammation, Barts Cancer Institute, Queen Mary University of London, Charterhouse Square, London EC1M 6BQ, UK. Electronic address: r.delaine
  • Burney S; School of Engineering and Materials Science, Institute of Bioengineering, Queen Mary University of London, Mile End, London E1 4NS, UK.
  • Balkwill FR; Centre for Cancer and Inflammation, Barts Cancer Institute, Queen Mary University of London, Charterhouse Square, London EC1M 6BQ, UK.
  • Knight MM; School of Engineering and Materials Science, Institute of Bioengineering, Queen Mary University of London, Mile End, London E1 4NS, UK.
J Mech Behav Biomed Mater ; 60: 401-415, 2016 07.
Article en En | MEDLINE | ID: mdl-26974584
ABSTRACT
Mechanical characterisation of soft biological tissues using standard compression or tensile testing presents a significant challenge due to specimen geometrical irregularities, difficulties in cutting intact and appropriately sized test samples, and issues with slippage or damage at the grips. Indentation can overcome these problems but requires fitting a model to the resulting load-displacement data in order to calculate moduli. Despite the widespread use of this technique, few studies experimentally validate their chosen model or compensate for boundary effects. In this study, viscoelastic hydrogels of different concentrations and dimensions were used to calibrate an indentation technique performed at large specimen-strain deformation (20%) and analysed with a range of routinely used mathematical models. A rigid, flat-ended cylindrical indenter was applied to each specimen from which 'indentation moduli' and relaxation properties were calculated and compared against values obtained from unconfined compression. Only one indentation model showed good agreement (<10% difference) with all moduli values obtained from compression. A sample thickness to indenter diameter ratio ≥11 and sample diameter to indenter diameter ratio ≥41 was necessary to achieve the greatest accuracy. However, it is not always possible to use biological samples within these limits, therefore we developed a series of correction factors. The approach was validated using human diseased omentum and bovine articular cartilage resulting in mechanical properties closely matching compression values. We therefore present a widely useable indentation analysis method to allow more accurate calculation of material mechanics which is important in the study of soft tissue development, ageing, health and disease.