Effect of calculation method on kidney dosimetry in 177Lu-octreotate treatment.
Heikkonen, Jorma; Mäenpää, Hanna; Hippeläinen, Eero; Reijonen, Vappu; Tenhunen, Mikko.
  • Heikkonen J; a Nuclear Medicine , Helsinki University Central Hospital, Cancer Center , Helsinki , Finland.
  • Mäenpää H; a Nuclear Medicine , Helsinki University Central Hospital, Cancer Center , Helsinki , Finland.
  • Hippeläinen E; b HUS Medical Imaging Center , Helsinki University Central Hospital, Helsinki , Finland.
  • Reijonen V; a Nuclear Medicine , Helsinki University Central Hospital, Cancer Center , Helsinki , Finland.
  • Tenhunen M; c Department of Radiotherapy , Helsinki University Central Hospital, Cancer Center , Helsinki , Finland.
Acta Oncol ; 55(9-10): 1069-1076, 2016.
Article en En | MEDLINE | ID: mdl-27219529
ABSTRACT

BACKGROUND:

177Lu-octreotate is an effective treatment modality for patients with metastatic neuroendocrine tumors. The kidney is a critical dose-limiting organ in that modality. We investigated the absorbed doses in the kidney and compared whole kidney volume (WKV) and small (4 cm3) volume of the kidney (SV) methods. We also evaluated a new calculation method that was based on two single photon emission computed tomography/computed tomography (SPECT/CT) scans.

METHODS:

Absorbed radiation doses in the kidneys were calculated for 24 patients with neuroendocrine tumors. All patients received four cycles of 177Lu-octreotate given at eight-week intervals with a mean activity of 7.1 GBq (range 3.28-8.79 GBq). Absorbed doses and half-lives were calculated by the WKV and SV methods. Dosimetry was determined for the cortex and medulla in the first treatment cycle.

RESULTS:

The mean absorbed radiation dose was 0.44 ± 0.15 Gy/GBq for the WKV method and, 0.74 ± 0.28 Gy/GBq for the SV method. Three patients had a 20% increase of the absorbed dose over the four treatment cycles for the WKV method compared to eight patients for the SV method. The mean absorbed dose in the medulla was 0.62 ± 0.27 Gy/GBq, whereas the mean absorbed dose in the cortex was 0.41 ± 0.22 Gy/GBq. Both regions had similar half-lives. Patients who received lower activities for medical reasons still had similar absorbed doses to kidneys compared to those who received the full activities. Our study indicates that absorbed doses can be calculated reliably using two SPECT/CT scans, at 24 and 168 hours after each treatment.

CONCLUSIONS:

Absorbed doses in the kidneys from systemic radionuclide therapy that are measured by the WKV method and SV method cannot be directly compared. There were regional differences within kidneys for the uptake of 177Lu-octreotate. Two SPECT/CTs are sufficient for kidney dosimetry based on our new calculation method.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Radioisótopos / Planificación de la Radioterapia Asistida por Computador / Octreótido / Tumores Neuroendocrinos / Riñón Tipo de estudio: Evaluation_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Año: 2016 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Radioisótopos / Planificación de la Radioterapia Asistida por Computador / Octreótido / Tumores Neuroendocrinos / Riñón Tipo de estudio: Evaluation_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Año: 2016 Tipo del documento: Article