Calibration of the OHREX high-resolution imaging crystal spectrometer at the Livermore electron beam ion traps.
Rev Sci Instrum
; 87(11): 11D604, 2016 Nov.
Article
en En
| MEDLINE
| ID: mdl-27910351
We report the calibration of the Orion High-Resolution X-ray (OHREX) imaging crystal spectrometer at the EBIT-I electron beam ion trap at Livermore. Two such instruments, dubbed OHREX-1 and OHREX-2, are fielded for plasma diagnostics at the Orion laser facility in the United Kingdom. The OHREX spectrometer can simultaneously house two spherically bent crystals with a radius of curvature of r = 67.2 cm. The focusing properties of the spectrometer allow both for larger distance to the source due to the increase in collected light and for observation of extended sources. OHREX is designed to cover a 2.5°-3° spectral range at Bragg angles around 51.3°. The typically high resolving powers at these large Bragg angles are ideally suited for line shape diagnostics. For instance, the nominal resolving power of the instrument (>10 000) is much higher than the effective resolving power associated with the Doppler broadening due to the temperature of the trapped ions in EBIT-I. The effective resolving power is only around 3000 at typical EBIT-I conditions, which nevertheless is sufficient to set up and test the instrument's spectral characteristics. We have calibrated the spectral range for a number of crystals using well known reference lines in the first and second order and derived the ion temperatures from these lines. We have also made use of the 50 µm size of the EBIT-I source width to characterize the spatial focusing of the spectrometer.
Search on Google
Banco de datos:
MEDLINE
Tipo de estudio:
Prognostic_studies
Idioma:
En
Año:
2016
Tipo del documento:
Article