Your browser doesn't support javascript.
loading
Replacing the Draize eye test: Impedance spectroscopy as a 3R method to discriminate between all GHS categories for eye irritation.
Lotz, C; Kiesewetter, L; Schmid, F F; Hansmann, J; Walles, H; Groeber-Becker, F.
  • Lotz C; Department Tissue Engineering & Regenerative Medicine (TERM), University Hospital Würzburg, Würzburg, 97070, Germany. Christian.Lotz@uni-wuerzburg.de.
  • Kiesewetter L; Department Tissue Engineering & Regenerative Medicine (TERM), University Hospital Würzburg, Würzburg, 97070, Germany.
  • Schmid FF; Translational Center for Regenerative Therapies, Fraunhofer Institute for Silicate Research, Würzburg, 97070, Germany.
  • Hansmann J; Department Tissue Engineering & Regenerative Medicine (TERM), University Hospital Würzburg, Würzburg, 97070, Germany.
  • Walles H; Translational Center for Regenerative Therapies, Fraunhofer Institute for Silicate Research, Würzburg, 97070, Germany.
  • Groeber-Becker F; Department Tissue Engineering & Regenerative Medicine (TERM), University Hospital Würzburg, Würzburg, 97070, Germany.
Sci Rep ; 8(1): 15049, 2018 10 09.
Article en En | MEDLINE | ID: mdl-30301970
ABSTRACT
Highly invasive animal based test procedures for risk assessment such as the Draize eye test are under increasing criticism due to poor transferability for the human organism and animal-welfare concerns. However, besides all efforts, the Draize eye test is still not completely replaced by alternative animal-free methods. To develop an in vitro test to identify all categories of eye irritation, we combined organotypic cornea models based on primary human cells with an electrical readout system that measures the impedance of the test models. First, we showed that employing a primary human cornea epithelial cell based model is advantageous in native marker expression to the primary human epidermal keratinocytes derived models. Secondly, by employing a non-destructive measuring system based on impedance spectroscopy, we could increase the sensitivity of the test system. Thereby, all globally harmonized systems categories of eye irritation could be identified by repeated measurements over a period of 7 days. Based on a novel prediction model we achieved an accuracy of 78% with a reproducibility of 88.9% to determine all three categories of eye irritation in one single test. This could pave the way according to the 3R principle to replace the Draize eye test.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Córnea / Ojo Tipo de estudio: Prognostic_studies / Risk_factors_studies Límite: Animals / Humans Idioma: En Año: 2018 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Córnea / Ojo Tipo de estudio: Prognostic_studies / Risk_factors_studies Límite: Animals / Humans Idioma: En Año: 2018 Tipo del documento: Article