Your browser doesn't support javascript.
loading
Dual-Energy Computed Tomography in Loosening of Revision Hip Prosthesis: A Comparison Between MARS and non-MARS Images.
Guzinski, Maciej; Kubicki, Konrad; Waszczuk, Lukasz; Morawska-Kochman, Monika; Kochman, Andrzej; Sasiadek, Marek.
  • Morawska-Kochman M; Department of Orthopedic and Traumatology, The Ministry of Internal Affairs Hospital, Wroclaw, Poland.
  • Kochman A; Department of Orthopedic and Traumatology, The Ministry of Internal Affairs Hospital, Wroclaw, Poland.
J Comput Assist Tomogr ; 43(3): 379-385, 2019.
Article en En | MEDLINE | ID: mdl-30855312
ABSTRACT

PURPOSE:

The aim of the study was to assess potential endoprosthesis loosening in patients after revision total hip arthroplasty, based on monochromatic dual-energy computed tomography (DECT) images obtained with and without metal artifact reduction software (MARS) and comparison with the clinical examination and hip function. MATERIALS AND

METHODS:

Twenty-five consecutive patients underwent DECT examinations. Two monochromatic data sets were generated with MARS (75-keV MARS) and without MARS (140-keV non-MARS) and evaluated for signs of loosening, using a dedicated radiological score. The Harris Hip Score (HHS) was used to evaluate patients' hip function.

RESULTS:

The mean radiological score for loosening in the non-MARS group was 16.2, in the MARS group 17.0 and was significantly higher (P < 0.001). Radiological loosening evaluation in non-MARS images correlated with the HHS score (ρ = 0.43, P = 0.03), whereas there was no correlation between MARS images and HHS (ρ = 0.15, P = 0.47). New MARS-related artifacts, which hinder implant-bone interface evaluation, were found in 75-keV MARS images ("pseudo-loosening" in 33 of 34 prostheses, "ground-glass blurring" in 20 of 32).

CONCLUSIONS:

High-energy monochromatic DECT images are superior to low-energy DECT MARS images in assessment of the potential loosening and correlate better with the clinical examination outcomes. For revision total hip arthroplasty evaluation, 2 data sets should be assessed inseparably because of extra artifacts in MARS images.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Reoperación / Falla de Prótesis / Artroplastia de Reemplazo de Cadera / Cadera / Prótesis de Cadera Tipo de estudio: Diagnostic_studies Límite: Female / Humans / Male / Middle aged Idioma: En Año: 2019 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Reoperación / Falla de Prótesis / Artroplastia de Reemplazo de Cadera / Cadera / Prótesis de Cadera Tipo de estudio: Diagnostic_studies Límite: Female / Humans / Male / Middle aged Idioma: En Año: 2019 Tipo del documento: Article