Ethanol fixation method for heart and lung imaging in micro-CT.
Patzelt, Matej; Mrzilkova, Jana; Dudak, Jan; Krejci, Frantisek; Zemlicka, Jan; Karch, Jakub; Musil, Vladimir; Rosina, Jozef; Sykora, Viktor; Horehledova, Barbora; Zach, Petr.
  • Patzelt M; Specialized Laboratory of Experimental Imaging, Third Faculty of Medicine, Charles University, Prague, Czech Republic.
  • Mrzilkova J; Institute of Experimental and Applied Physics and Faculty of Biomedical Engineering, Czech Technical University in Prague, Prague, Czech Republic.
  • Dudak J; Department of Anatomy, Third Faculty of Medicine, Charles University, Ruska 87, 100 34, Prague, Czech Republic.
  • Krejci F; Specialized Laboratory of Experimental Imaging, Third Faculty of Medicine, Charles University, Prague, Czech Republic. jana.mrzilkova@lf3.cuni.cz.
  • Zemlicka J; Institute of Experimental and Applied Physics and Faculty of Biomedical Engineering, Czech Technical University in Prague, Prague, Czech Republic. jana.mrzilkova@lf3.cuni.cz.
  • Karch J; Department of Anatomy, Third Faculty of Medicine, Charles University, Ruska 87, 100 34, Prague, Czech Republic. jana.mrzilkova@lf3.cuni.cz.
  • Musil V; Specialized Laboratory of Experimental Imaging, Third Faculty of Medicine, Charles University, Prague, Czech Republic.
  • Rosina J; Institute of Experimental and Applied Physics and Faculty of Biomedical Engineering, Czech Technical University in Prague, Prague, Czech Republic.
  • Sykora V; Institute of Experimental and Applied Physics, Czech Technical University in Prague, Prague, Czech Republic.
  • Horehledova B; Faculty of Biomedical Engineering, Czech Technical University in Prague, Kladno, Czech Republic.
  • Zach P; Specialized Laboratory of Experimental Imaging, Third Faculty of Medicine, Charles University, Prague, Czech Republic.
Jpn J Radiol ; 37(6): 500-510, 2019 Jun.
Article en En | MEDLINE | ID: mdl-30875011
ABSTRACT

PURPOSE:

The soft tissue imaging in micro-CT remains challenging due to its low intrinsic contrast. The aim of this study was to create a simple staining method omitting the usage of contrast agents for ex vivo soft tissue imaging in micro-CT. MATERIALS AND

METHODS:

Hearts and lungs from 30 mice were used. Twenty-seven organs were either fixed in 97% or 50% ethanol solution or in a series of ascending ethanol concentrations. Images were acquired after 72, 168 and 336 h on a custom-built micro-CT machine and compared to scans of three native samples.

RESULTS:

Ethanol provided contrast enhancement in all evaluated fixations. Fixation in 97% ethanol resulted in contrast enhancement after 72 h; however, it caused hardening of the samples. Fixation in 50% ethanol provided contrast enhancement after 336 h, with milder hardening, compared to the 97% ethanol fixation, but the visualization of details was worse. The fixation in a series of ascending ethanol concentrations provided the most satisfactory results; all organs were visualized in great detail without tissue damage.

CONCLUSIONS:

Simple ethanol fixation improves the tissue contrast enhancement in micro-CT. The best results can be obtained with fixation of the soft tissue samples in a series of ascending ethanol concentrations.