Effect of respiratory motion on free-breathing 3D stack-of-radial liver R 2 ∗ relaxometry and improved quantification accuracy using self-gating.
Magn Reson Med
; 83(6): 1964-1978, 2020 06.
Article
en En
| MEDLINE
| ID: mdl-31682016
ABSTRACT
PURPOSE:
To develop an accurate free-breathing 3D liver R2∗ mapping approach and to evaluate it in vivo.METHODS:
A free-breathing multi-echo stack-of-radial sequence was applied in 5 normal subjects and 6 patients at 3 Tesla. Respiratory motion compensation was implemented using the inherent self-gating signal. A breath-hold Cartesian acquisition was the reference standard. Proton density fat fraction and R2∗ were measured and compared between radial and Cartesian methods using Bland-Altman plots. The normal subject results were fitted to a linear mixed model (P < .05 considered significant).RESULTS:
Free-breathing stack-of-radial without self-gating exhibited signal attenuation in echo images and artifactually elevated apparent R2∗ values. In the Bland-Altman plots of normal subjects, compared to breath-hold Cartesian, free-breathing stack-of-radial acquisitions of 22, 30, 36, and 44 slices, had mean R2∗ differences of 27.4, 19.4, 10.9, and 14.7 s-1 with 800 radial views, and they had 18.4, 11.9, 9.7, and 27.7 s-1 with 404 views, which were reduced to 0.4, 0.9, -0.2, and -0.7 s-1 and to -1.7, -1.9, -2.1, and 0.5 s-1 with self-gating, respectively. No substantial proton density fat fraction differences were found. The linear mixed model showed free-breathing radial R2∗ results without self-gating were significantly biased by 17.2 s-1 averagely (P = .002), which was eliminated with self-gating (P = .930). Proton density fat fraction results were not different (P > .234). For patients, Bland-Altman plots exhibited mean R2∗ differences of 14.4 and 0.1 s-1 for free-breathing stack-of-radial without self-gating and with self-gating, respectively, but no substantial proton density fat fraction differences.CONCLUSION:
The proposed self-gating method corrects the respiratory motion bias and enables accurate free-breathing stack-of-radial quantification of liver R2∗ .Palabras clave
Texto completo:
1
Banco de datos:
MEDLINE
Asunto principal:
Interpretación de Imagen Asistida por Computador
/
Imagenología Tridimensional
Límite:
Humans
Idioma:
En
Año:
2020
Tipo del documento:
Article