Effect of respiratory motion on free-breathing 3D stack-of-radial liver R2 relaxometry and improved quantification accuracy using self-gating.
Zhong, Xiaodong; Armstrong, Tess; Nickel, Marcel D; Kannengiesser, Stephan A R; Pan, Li; Dale, Brian M; Deshpande, Vibhas; Kiefer, Berthold; Wu, Holden H.
  • Zhong X; MR R&D Collaborations, Siemens Healthcare, Los Angeles, California.
  • Armstrong T; Department of Radiological Sciences, David Geffen School of Medicine, University of California Los Angeles, Los Angeles, California.
  • Nickel MD; Physics and Biology in Medicine Interdepartmental Program, David Geffen School of Medicine, University of California Los Angeles, Los Angeles, California.
  • Kannengiesser SAR; MR Application Development, Siemens Healthcare GmbH, Erlangen, Germany.
  • Pan L; MR Application Development, Siemens Healthcare GmbH, Erlangen, Germany.
  • Dale BM; MR R&D Collaborations, Siemens Healthcare, Baltimore, Maryland.
  • Deshpande V; MR R&D Collaborations, Siemens Healthcare, Cary, North Carolina.
  • Kiefer B; MR R&D Collaborations, Siemens Healthcare, Austin, Texas.
  • Wu HH; MR Application Development, Siemens Healthcare GmbH, Erlangen, Germany.
Magn Reson Med ; 83(6): 1964-1978, 2020 06.
Article en En | MEDLINE | ID: mdl-31682016
ABSTRACT

PURPOSE:

To develop an accurate free-breathing 3D liver R2∗ mapping approach and to evaluate it in vivo.

METHODS:

A free-breathing multi-echo stack-of-radial sequence was applied in 5 normal subjects and 6 patients at 3 Tesla. Respiratory motion compensation was implemented using the inherent self-gating signal. A breath-hold Cartesian acquisition was the reference standard. Proton density fat fraction and R2∗ were measured and compared between radial and Cartesian methods using Bland-Altman plots. The normal subject results were fitted to a linear mixed model (P < .05 considered significant).

RESULTS:

Free-breathing stack-of-radial without self-gating exhibited signal attenuation in echo images and artifactually elevated apparent R2∗ values. In the Bland-Altman plots of normal subjects, compared to breath-hold Cartesian, free-breathing stack-of-radial acquisitions of 22, 30, 36, and 44 slices, had mean R2∗ differences of 27.4, 19.4, 10.9, and 14.7 s-1 with 800 radial views, and they had 18.4, 11.9, 9.7, and 27.7 s-1 with 404 views, which were reduced to 0.4, 0.9, -0.2, and -0.7 s-1 and to -1.7, -1.9, -2.1, and 0.5 s-1 with self-gating, respectively. No substantial proton density fat fraction differences were found. The linear mixed model showed free-breathing radial R2∗ results without self-gating were significantly biased by 17.2 s-1 averagely (P = .002), which was eliminated with self-gating (P = .930). Proton density fat fraction results were not different (P > .234). For patients, Bland-Altman plots exhibited mean R2∗ differences of 14.4 and 0.1 s-1 for free-breathing stack-of-radial without self-gating and with self-gating, respectively, but no substantial proton density fat fraction differences.

CONCLUSION:

The proposed self-gating method corrects the respiratory motion bias and enables accurate free-breathing stack-of-radial quantification of liver R2∗ .