Bipolar Directional Deep Brain Stimulation in Essential and Parkinsonian Tremor.
Steffen, Julia K; Reker, Paul; Mennicken, Fiona K; Dembek, Till A; Dafsari, Haidar S; Fink, Gereon R; Visser-Vandewalle, Veerle; Barbe, Michael T.
  • Steffen JK; Faculty of Medicine and University Hospital Cologne, Department of Neurology, University of Cologne, Cologne, Germany.
  • Reker P; Faculty of Medicine and University Hospital Cologne, Department of Neurology, University of Cologne, Cologne, Germany.
  • Mennicken FK; Faculty of Medicine and University Hospital Cologne, Department of Neurology, University of Cologne, Cologne, Germany.
  • Dembek TA; Faculty of Medicine and University Hospital Cologne, Department of Neurology, University of Cologne, Cologne, Germany.
  • Dafsari HS; Faculty of Medicine and University Hospital Cologne, Department of Neurology, University of Cologne, Cologne, Germany.
  • Fink GR; Faculty of Medicine and University Hospital Cologne, Department of Neurology, University of Cologne, Cologne, Germany.
  • Visser-Vandewalle V; Cognitive Neuroscience, Institute of Neuroscience and Medicine (INM-3), Research Center Jülich, Jülich, Germany.
  • Barbe MT; Faculty of Medicine and University Hospital Cologne, Department of Stereotactic and Functional Neurosurgery, University of Cologne, Cologne, Germany.
Neuromodulation ; 23(4): 543-549, 2020 Jun.
Article en En | MEDLINE | ID: mdl-32040883
ABSTRACT

OBJECTIVE:

To compare directional monopolar, bipolar, and directional bipolar thalamic deep brain stimulation (DBS) in tremor patients.

METHODS:

Fourteen tremor patients (7 Essential Tremor and 7 Parkinson's Disease) implanted with directional DBS electrodes in the ventral intermediate nucleus (VIM) were enrolled. Side-effect thresholds of monopolar directional stimulation (DIRECT) were compared to circular DBS as well as, in a randomized design, to those of two different bipolar stimulation settings (BIPOLAR = circular anode; BI-DIRECT = directional anode). Tremor suppression (Tremor Rating Scale, TRS) right below the side-effect threshold was also assessed.

RESULTS:

Directional DBS in the individually best direction showed higher side-effect thresholds than circular DBS (p = 0.0063). The thresholds were raised further using either one of the bipolar stimulation paradigms (BIPOLAR p = 0.0029, BI-DIRECT p = 0.0022). The side-effect thresholds did not differ between both bipolar settings, but side-effects were less frequent with BI-DIRECT. No difference in TRS scores with stimulation just below the side-effect threshold was found between all stimulation conditions.

CONCLUSIONS:

Side-effect thresholds of monopolar directional and bipolar stimulation with both circular and directional anodes were higher compared to traditional monopolar circular stimulation in the VIM. Bipolar DBS with directional anodes evoked side-effect less frequently than bipolar and monopolar directional stimulation. All stimulation settings had comparable effects on tremor suppression just below their side-effect thresholds. Thus, directional and different bipolar settings should be explored in patients with bothersome side-effects of thalamic stimulation when monopolar stimulation settings are not satisfying. Further studies are needed to explore the efficiency of the different bipolar stimulation paradigms.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Enfermedad de Parkinson / Temblor Esencial / Estimulación Encefálica Profunda Tipo de estudio: Clinical_trials Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Año: 2020 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Enfermedad de Parkinson / Temblor Esencial / Estimulación Encefálica Profunda Tipo de estudio: Clinical_trials Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Año: 2020 Tipo del documento: Article