Your browser doesn't support javascript.
loading
Utilization of Novel High-Resolution, MRI-Based Vascular Imaging Modality for Preoperative Stereoelectroencephalography Planning in Children: A Technical Note.
Feng, Austin Y; Ho, Allen L; Kim, Lily H; Sussman, Eric S; Pendharkar, Arjun V; Iv, Michael; Yeom, Kristen W; Halpern, Casey H; Grant, Gerald A.
  • Feng AY; Department of Neurosurgery, Stanford University School of Medicine, Stanford, California, USA.
  • Ho AL; Department of Neurosurgery, Stanford University School of Medicine, Stanford, California, USA.
  • Kim LH; Department of Neurosurgery, Stanford University School of Medicine, Stanford, California, USA.
  • Sussman ES; Department of Neurosurgery, Stanford University School of Medicine, Stanford, California, USA.
  • Pendharkar AV; Department of Neurosurgery, Stanford University School of Medicine, Stanford, California, USA.
  • Iv M; Department of Radiology, Stanford University Medical Center, Stanford, California, USA.
  • Yeom KW; Department of Radiology, Pediatric Radiology, Lucile Packard Children's Hospital at Stanford, Stanford, California, USA.
  • Halpern CH; Department of Neurosurgery, Stanford University School of Medicine, Stanford, California, USA.
  • Grant GA; Department of Neurosurgery, Stanford University School of Medicine, Stanford, California, USA, ggrant2@Stanford.edu.
Stereotact Funct Neurosurg ; 98(1): 1-7, 2020.
Article en En | MEDLINE | ID: mdl-32062664
ABSTRACT

INTRODUCTION:

Stereoelectroencephalography (SEEG) is a powerful intracranial diagnostic tool that requires accurate imaging for proper electrode trajectory planning to ensure efficacy and maximize patient safety. Computed tomography (CT) angiography and digital subtraction angiography are commonly used, but recent developments in magnetic resonance angiography allow for high-resolution vascular visualization without added risks of radiation. We report on the accuracy of electrode placement under robotic assistance planning utilizing a novel high-resolution magnetic resonance imaging (MRI)-based imaging modality.

METHODS:

Sixteen pediatric patients between February 2014 and October 2017 underwent SEEG exploration for epileptogenic zone localization. A gadolinium-enhanced 3D T1-weighted spoiled gradient recalled echo sequence with minimum echo time and repetition time was applied for background parenchymal suppression and vascular enhancement. Electrode placement accuracy was determined by analyzing postoperative CT scans laid over preoperative virtual electrode trajectory paths. Entry point, target point, and closest vessel intersection were measured.

RESULTS:

For any intersection along the trajectory path, 57 intersected vessels were measured. The mean diameter of an intersected vessel was 1.0343 ± 0.1721 mm, and 21.05% of intersections involved superficial vessels. There were 157 overall intersection + near-miss events. The mean diameter for an involved vessel was 1.0236 ± 0.0928 mm, and superficial vessels were involved in 20.13%. Looking only at final electrode target, 3 intersection events were observed. The mean diameter of an intersected vessel was 1.0125 ± 0.2227 mm. For intersection + near-miss events, 24 were measured. An involved vessel's mean diameter was 1.1028 ± 0.2634 mm. For non-entry point intersections, 45 intersected vessels were measured. The mean diameter for intersected vessels was 0.9526 ± 0.0689 mm. For non-entry point intersections + near misses, 126 events were observed. The mean diameter for involved vessels was 0.9826 ± 0.1008 mm.

CONCLUSION:

We believe this novel sequence allows better identification of superficial and deeper subcortical vessels compared to conventional T1-weighted gadolinium-enhanced MRI.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Cuidados Preoperatorios / Imagen por Resonancia Magnética / Tomografía Computarizada por Rayos X / Técnicas Estereotáxicas / Electroencefalografía / Epilepsia Refractaria Límite: Adolescent / Child / Female / Humans / Male Idioma: En Año: 2020 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Cuidados Preoperatorios / Imagen por Resonancia Magnética / Tomografía Computarizada por Rayos X / Técnicas Estereotáxicas / Electroencefalografía / Epilepsia Refractaria Límite: Adolescent / Child / Female / Humans / Male Idioma: En Año: 2020 Tipo del documento: Article