Your browser doesn't support javascript.
loading
Reliability of EOS compared to conventional radiographs for evaluation of lower extremity deformity in adult patients.
Wise, Kelsey L; Kelly, Brandon J; Agel, Julie; Marette, Shelly; Macalena, Jeffrey A.
  • Wise KL; Department of Orthopedic Surgery, University of Minnesota, 2450 Riverside Avenue South, Suite R200, Minneapolis, MN, 55454, USA.
  • Kelly BJ; Department of Orthopedic Surgery, University of Minnesota, 2450 Riverside Avenue South, Suite R200, Minneapolis, MN, 55454, USA.
  • Agel J; Department of Orthopedic Surgery, University of Minnesota, 2450 Riverside Avenue South, Suite R200, Minneapolis, MN, 55454, USA.
  • Marette S; Department of Radiology, University of Minnesota, 420 Delaware St. SE, Minneapolis, MN, 55455, USA.
  • Macalena JA; Department of Orthopedic Surgery, University of Minnesota, 2450 Riverside Avenue South, Suite R200, Minneapolis, MN, 55454, USA. maca0049@umn.edu.
Skeletal Radiol ; 49(9): 1423-1430, 2020 Sep.
Article en En | MEDLINE | ID: mdl-32291475
ABSTRACT

OBJECTIVE:

The purpose of this study was to compare reliability of lower extremity imaging measurements using EOS and conventional X-ray (CR) of adult patients with mechanical axis malalignment. MATERIALS AND

METHODS:

Ten patients (20 lower limbs) of mean age of 31.6 years (range 21-39) with post-traumatic deformities who presented for evaluation of osteotomies and/or ligament and cartilage reconstructions were prospectively enrolled. Two independent observers performed full-length anterior-posterior (AP) measurements 2 weeks apart on both CXR and two-dimensional (2D) EOS images. Measurements included weight-bearing axis (WBA), varus/valgus angle (V/V), femoral length (FL), tibial length (TL), femoral mechanical axis (FMA), tibial mechanical axis (TMA), and total limb length (TLL). Reliability was determined with random effects modeling of intraclass correlation coefficients (ICC) set to consistency. Three statistical operations were performed to compare interrater validity in CXR and EOS students' two-sample t test, paired two-sample t test, and Pearson's correlative r-statistical agreement.

RESULTS:

There was a statistically significant difference for V/V, FL, and TLL (all p < 0.01) between CXR and EOS. A relatively large proportion of the population consistently had larger V/V measures for EOS compared to CXR. In contrast, the FL and TLL measures were consistently larger for CXR compared to EOS. The differences between CXR and EOS measurements were statistically significant, though the small differences in values were not clinically meaningful. Agreement of all measures remained high (r = 0.84-0.99).

CONCLUSION:

Using 2D EOS for lower extremity measurements is reproducible, reliable, and comparable to the gold standard, standing long leg radiographs.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Tibia / Fémur Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Adult / Humans Idioma: En Año: 2020 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Tibia / Fémur Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Adult / Humans Idioma: En Año: 2020 Tipo del documento: Article