Is signal resolution of linear cryocatheters appropriate to map superoparaseptal accessory pathways? A comparison with Orion high-resolution mapping.
Ducceschi, Valentino; Maddaluno, Francesco; Del Giorno, Giuseppe; Catalano, Angelo; Carbone, Angelo.
  • Ducceschi V; Pellegrini Hospital Napoli Italy.
  • Maddaluno F; Boston Scientific Milan Italy.
  • Del Giorno G; Maria Santissima Addolorata Hospital Eboli (SA) Italy.
  • Catalano A; Maria Santissima Addolorata Hospital Eboli (SA) Italy.
  • Carbone A; Maria Santissima Addolorata Hospital Eboli (SA) Italy.
J Arrhythm ; 36(5): 942-944, 2020 Oct.
Article en En | MEDLINE | ID: mdl-33024475
ABSTRACT
Larger tip cryoablation catheters can dampen the signal at the distal tip reducing the resolution of mapping of supero paraseptal APs. The aid of high resolution mapping should be considered to improve safety and efficacy.