Operative Complications of Glaucoma Drainage Implant Tube Insertion Through the Sulcus for Pseudophakic Eye.
Asaoka, Satoko; Kasuga, Toshimitsu; Matsunaga, Toru; Hayashi, Yusuke; Asada, Yosuke; Iwamoto, Satoshi; Hirakata, Toshiaki; Honda, Rio; Obazawa, Hanako; Sasaki, Hidenori; Ohta, Toshihiko; Matsuda, Akira.
  • Asaoka S; Department of Ophthalmology, Juntendo University School of Medicine.
  • Kasuga T; Department of Ophthalmology, Juntendo University Shizuoka Hospital, Izunokuni.
  • Matsunaga T; Department of Ophthalmology, Juntendo University School of Medicine.
  • Hayashi Y; SEED Co. Ltd, Tokyo, Japan.
  • Asada Y; Department of Ophthalmology, Juntendo University School of Medicine.
  • Iwamoto S; Department of Ophthalmology, Juntendo University Shizuoka Hospital, Izunokuni.
  • Hirakata T; Department of Ophthalmology, Juntendo University School of Medicine.
  • Honda R; Department of Ophthalmology, Juntendo University School of Medicine.
  • Obazawa H; Department of Ophthalmology, Juntendo University School of Medicine.
  • Sasaki H; Department of Ophthalmology, Juntendo University School of Medicine.
  • Ohta T; Department of Ophthalmology, Koshigaya Municipal Hospital, Koshigaya.
  • Matsuda A; Department of Ophthalmology, Asama General Hospital, Saku, Japan.
J Glaucoma ; 30(4): e169-e174, 2021 04 01.
Article en En | MEDLINE | ID: mdl-33449583
ABSTRACT
PRECIS Malposition of the tube through the ciliary sulcus is more frequently observed with the Ahmed glaucoma valve (AGV) than the Baerveldt drainage implant (BDI) due to the weaker rigidity of the Ahmed tube.

PURPOSE:

To report intraoperative and early postoperative complications of ciliary sulcus tube insertion of glaucoma drainage implants (GDIs). PATIENTS AND

METHODS:

We performed retrospective analysis of 104 eyes of 94 patients with GDI tube insertion through the ciliary sulcus were performed. The rigidities of tubes were also examined using a microcompression tester.

RESULTS:

The mean observation period was 20.0 (range, 6 to 60) months. Thirteen eyes were treated with the BDI and 91 were with the AGV. The mean age of the patients was 69.3 (34 to 90) years. The mean intraocular pressure was 27.9 mm Hg before surgery and 12.9 mm Hg after surgery (P<0.01). Upon tube insertion 42/91 eyes (46%) with the AGV required reinsertion of the tube due to malpositioning, whereas only 1/13 (8%) eyes with BDI did (P<0.01). Transient hyphema (12 eyes) and hypotony (12 eyes) were observed as early postoperative complications with the AGV. Seven eyes with hypotony were treated by proline stenting of the tube. We could not accomplish sulcus insertions in 4 eyes. Microcompression analysis of the tubes showed that the BGI tube was more rigid than that of the AGV.

CONCLUSIONS:

Ciliary sulcus insertion of the tube is an effective method to control intraocular pressure. The tube of the AGV was more difficult to insert through the sulcus than the BDI due to its weaker rigidity.