Your browser doesn't support javascript.
loading
[Bileaflet mechanical valve dysfunction related to fibrous pannus: the role of cine-fluoroscopy]. / Disfunzione di protesi valvolare meccanica bileaflet da panno fibroso: non dimenticare la cine-fluoroscopia!
Troccolo, Alessandro; De Marzo, Vincenzo; Della Bona, Roberta; Seitun, Sara; Santini, Francesco; Porto, Italo.
  • Troccolo A; Divisione di Cardiologia, DICATOV-Dipartimento Cardio-Toraco-Vascolare, Ospedale Policlinico San Martino IRCCS per l'Oncologia, Genova - Scuola di Scienze Mediche e Farmaceutiche, Università degli Studi, Policlinico San Martino IRCCS, Genova.
  • De Marzo V; Divisione di Cardiologia, DICATOV-Dipartimento Cardio-Toraco-Vascolare, Ospedale Policlinico San Martino IRCCS per l'Oncologia, Genova - Scuola di Scienze Mediche e Farmaceutiche, Università degli Studi, Policlinico San Martino IRCCS, Genova.
  • Della Bona R; Divisione di Cardiologia, DICATOV-Dipartimento Cardio-Toraco-Vascolare, Ospedale Policlinico San Martino IRCCS per l'Oncologia, Genova - Scuola di Scienze Mediche e Farmaceutiche, Università degli Studi, Policlinico San Martino IRCCS, Genova.
  • Seitun S; Dipartimento di Radiodiagnostica, Ospedale Policlinico San Martino IRCCS per l'Oncologia, Genova.
  • Santini F; Divisione di Cardiochirurgia, DICATOV-CardioThoracic and Vascular Department, Ospedale Policlinico San Martino IRCCS per l'Oncologia, Genova.
  • Porto I; Divisione di Cardiologia, DICATOV-Dipartimento Cardio-Toraco-Vascolare, Ospedale Policlinico San Martino IRCCS per l'Oncologia, Genova - Scuola di Scienze Mediche e Farmaceutiche, Università degli Studi, Policlinico San Martino IRCCS, Genova.
G Ital Cardiol (Rome) ; 22(6): 490-493, 2021 Jun.
Article en It | MEDLINE | ID: mdl-34037593
ABSTRACT
We report the case of a 72-year-old female patient admitted for worsening heart failure. The patient had undergone aortic valve replacement with a mechanical prosthesis 28 years before and since then she had several acute heart failure episodes, with a progressive increase in transprosthetic gradients, without identifying a specific cause.We describe the diagnostic tools used to reach a diagnosis, with particular emphasis on the use of cine-fluoroscopy that allowed to make the decisively diagnosis of prosthetic valve dysfunction, subsequently confirmed by cardiac computed tomography (CT). By cine-fluoroscopy, a widespread, easy, low-cost, and safe tool (no need for medium contrast and low radiation dose), it is possible to precisely define the function of the valve leaflets and measure their opening and closing angles, comparing them to the specific reference angles. To make the correct diagnosis we also performed a cardiac CT demonstrating a sub-aortic fibrous pannus. However, although cardiac CT is highly accurate for the identification of valve leaflet neoformations and abnormalities and for the discrimination between thrombotic formations and fibrous pannus, it is burdened by high costs, use of contrast medium, and limited available dedicated devices.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Trombosis / Prótesis Valvulares Cardíacas Tipo de estudio: Prognostic_studies Límite: Aged / Female / Humans Idioma: It Año: 2021 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Trombosis / Prótesis Valvulares Cardíacas Tipo de estudio: Prognostic_studies Límite: Aged / Female / Humans Idioma: It Año: 2021 Tipo del documento: Article