Your browser doesn't support javascript.
loading
The use of ANOVA-PCA and DD-SIMCA in the development of corn flour laboratory reference materials.
Dos Santos, Liz O; Dos Santos, Ana M P; Ferreira, Márcia M C; Ferreira, Sergio L C; Nepomuceno, Ana Flávia S F.
  • Dos Santos LO; Universidade Federal da Bahia, Instituto de Química, Grupo de Pesquisa em Química e Quimiometria, Campus Ondina, 41170-115 Salvador, Bahia, Brazil; Instituto Nacional de Ciência e Tecnologia, INCT, de Energia e Ambiente, Universidade Federal da Bahia, 40170-290 Salvador, Bahia, Brazil; Universidade
  • Dos Santos AMP; Universidade Federal da Bahia, Instituto de Química, Grupo de Pesquisa em Química e Quimiometria, Campus Ondina, 41170-115 Salvador, Bahia, Brazil; Instituto Nacional de Ciência e Tecnologia, INCT, de Energia e Ambiente, Universidade Federal da Bahia, 40170-290 Salvador, Bahia, Brazil.
  • Ferreira MMC; Theoretical and Applied Chemometrics Laboratory (LQTA), Institute of Chemistry, University of Campinas-Unicamp, P.O. Box 6154, Campinas, SP 13084-971, Brazil.
  • Ferreira SLC; Universidade Federal da Bahia, Instituto de Química, Grupo de Pesquisa em Química e Quimiometria, Campus Ondina, 41170-115 Salvador, Bahia, Brazil; Instituto Nacional de Ciência e Tecnologia, INCT, de Energia e Ambiente, Universidade Federal da Bahia, 40170-290 Salvador, Bahia, Brazil.
  • Nepomuceno AFSF; Universidade Federal da Bahia, Instituto de Química, Grupo de Pesquisa em Química e Quimiometria, Campus Ondina, 41170-115 Salvador, Bahia, Brazil.
Food Chem ; 367: 130748, 2022 Jan 15.
Article en En | MEDLINE | ID: mdl-34375894
The development of a collaborative study as a requirement for the preparation of a laboratory reference material candidate is reported in this paper. The evaluation was performed by 13 laboratories invited to quantify the calcium, potassium, magnesium, phosphorus, copper, iron, manganese and zinc; 8 of them presented results for all the analytes under investigation. The data were statistically analyzed by applying the z-score robust technique as recommended by ISO Guide 35. For the potassium element, laboratories 4 and 13 presented questionable results. Laboratory 5 proved to be unsatisfactory for calcium and zinc. ANOVA-PCA and DD-SIMCA were also applied to evaluate stability and interlaboratory studies results, respectively. It has been demonstrated that multivariate data analysis can be successfully applied as an alternative method to the recommendations made by ISO 13528 and ISO Guide 35 with defined confidence intervals.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Zea mays / Laboratorios Idioma: En Año: 2022 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Zea mays / Laboratorios Idioma: En Año: 2022 Tipo del documento: Article