Your browser doesn't support javascript.
loading
Comparison of Ustekinumab Trough Concentrations Measured by 2 ELISA Kits and Evaluation of Clinical Response in Crohn's Disease.
Kwon, Yiyoung; Kang, Ben; Kim, Eun Sil; Choe, Yon Ho; Kim, Mi Jin.
  • Kwon Y; Department of Pediatrics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea; and.
  • Kang B; Department of Pediatrics, School of Medicine, Kyungpook National University Children's Hospital, Kyungpook National University, Daegu, Korea.
  • Kim ES; Department of Pediatrics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea; and.
  • Choe YH; Department of Pediatrics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea; and.
  • Kim MJ; Department of Pediatrics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea; and.
Ther Drug Monit ; 44(4): 535-542, 2022 08 01.
Article en En | MEDLINE | ID: mdl-35175246
ABSTRACT

BACKGROUND:

Ustekinumab is a recently introduced biological agent for the treatment of Crohn's disease. The clinical use of the trough concentration of ustekinumab is not as standardized as that of infliximab. The authors aimed to introduce a measurement method and the results of trough concentrations of ustekinumab in clinical applications.

METHODS:

Thirty-two blood samples from 10 young adult patients diagnosed with Crohn's disease were analyzed. During the maintenance treatment, injection intervals were shortened from 12 weeks to 8 weeks in 4 patients who exhibited a loss of response. Ustekinumab trough concentrations were measured using 2 commercial ELISA kits, kit A and kit B.

RESULTS:

The median trough concentrations measured with kits A and B were 0.26 and 0.38 mcg/mL, respectively. In the case of kit A, low trough concentrations were undetected on many occasions and measured as zero, whereas kit B displayed their relative values even at low concentrations. Poor clinical parameters, elevated erythrocyte sedimentation rate, C-reactive protein, and calprotectin levels were significantly correlated with lower trough concentrations ( P < 0.05). The area under the receiver operating characteristics curve of kit B (0.921) was greater than that of kit A (0.744). The optimal cutoff values for prediction clinical responses were 0.17 and 0.41 mcg/mL for kit A and kit B, respectively.

CONCLUSIONS:

The trough concentration of ustekinumab measured by the 2 ELISA kits correlated with laboratory results that indicated the activity of Crohn's disease. Furthermore, kit B detected even minute changes in trough concentrations.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Enfermedad de Crohn / Ustekinumab Tipo de estudio: Prognostic_studies Límite: Adult / Humans Idioma: En Año: 2022 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Enfermedad de Crohn / Ustekinumab Tipo de estudio: Prognostic_studies Límite: Adult / Humans Idioma: En Año: 2022 Tipo del documento: Article