Your browser doesn't support javascript.
loading
Mean glandular dose in the mammary gland and dose of radiation in the thyroid gland and lens in women with and without breast implants during different modalities of mammography.
Pérez Fuentes, J A; Roldán Sánchez, V S; Gordillo Ledesma, A K; Mena, A F; Brito, S; Soteldo, C.
  • Pérez Fuentes JA; Centro Clínico de Estereotaxia - CECLINES, Caracas, Venezuela. Electronic address: perezfuentesjorge@gmail.com.
  • Roldán Sánchez VS; Instituto Venezolano de Investigaciones Científicas, Edo, Miranda, Venezuela.
  • Gordillo Ledesma AK; Centro Clínico de Estereotaxia - CECLINES, Caracas, Venezuela.
  • Mena AF; Centro Clínico de Estereotaxia - CECLINES, Caracas, Venezuela.
  • Brito S; Centro Clínico de Estereotaxia - CECLINES, Caracas, Venezuela.
  • Soteldo C; Instituto Venezolano de Investigaciones Científicas, Edo, Miranda, Venezuela.
Radiologia (Engl Ed) ; 64 Suppl 1: 11-19, 2022 Mar.
Article en En | MEDLINE | ID: mdl-35428461
ABSTRACT
There have always been concerns about the secondary effects of diagnostic methods that use ionizing radiation. During mammography, the parameters to be concerned about are the mean glandular dose and the scatter dose. We evaluated the dose of radiation to the breast, thyroid gland, and lens in digital mammography in women with and without implants, in tomosynthesis in women with and without implants, and in contrast-enhanced mammography. MATERIALS AND

METHODS:

The study included 212 women with and without disease who were attended at the Centro Clínico de Estereotaxia, CECLINES, in Caracas, Venezuela, between June 2017 and August 2017; the women were classified into five groups according to the mammographic modality used to evaluate them and whether or not they had implants. The statistical analysis included descriptive statistics for the study population. We used the Mann-Whitney U to compare the mean glandular dose and dose in the thyroid gland and lens between groups.

RESULTS:

The mean glandular dose and the dose of radiation received in the thyroid and lens were within the acceptable range. In a few exceptions, the mean glandular dose per view was slightly higher than 3 mGy. The scatter dose to the thyroid gland and the lens during mammography has a very small contribution to the annual dose equivalent.

CONCLUSION:

The mean glandular dose and the scatter dose to the thyroid gland and lens delivered during tomosynthesis and 2D mammography in women with implants were higher than those delivered during other mammographic techniques in women without implants.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Mamografía / Implantes de Mama / Glándulas Mamarias Humanas Límite: Female / Humans / Male Idioma: En Año: 2022 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Mamografía / Implantes de Mama / Glándulas Mamarias Humanas Límite: Female / Humans / Male Idioma: En Año: 2022 Tipo del documento: Article