Three-Dimensional Near-Infrared Specimen Mapping Can Identify the Distance from the Tumor to the Surgical Margin During Resection of Pulmonary Ground Glass Opacities.
Kennedy, Gregory T; Azari, Feredun S; Bernstein, Elizabeth; Deshpande, Charuhas; Kucharczuk, John C; Delikatny, Edward J; Singhal, Sunil.
  • Kennedy GT; Department of Surgery, University of Pennsylvania School of Medicine, 3400 Spruce Street, 6 White Building, Philadelphia, PA, 19104, USA.
  • Azari FS; Department of Surgery, University of Pennsylvania School of Medicine, 3400 Spruce Street, 6 White Building, Philadelphia, PA, 19104, USA.
  • Bernstein E; Department of Surgery, University of Pennsylvania School of Medicine, 3400 Spruce Street, 6 White Building, Philadelphia, PA, 19104, USA.
  • Deshpande C; Department of Pathology, University of Pennsylvania School of Medicine, Philadelphia, PA, USA.
  • Kucharczuk JC; Department of Surgery, University of Pennsylvania School of Medicine, 3400 Spruce Street, 6 White Building, Philadelphia, PA, 19104, USA.
  • Delikatny EJ; Department of Radiology, University of Pennsylvania School of Medicine, Philadelphia, PA, USA.
  • Singhal S; Department of Surgery, University of Pennsylvania School of Medicine, 3400 Spruce Street, 6 White Building, Philadelphia, PA, 19104, USA. Sunil.Singhal@pennmedicine.upenn.edu.
Mol Imaging Biol ; 25(1): 203-211, 2023 02.
Article en En | MEDLINE | ID: mdl-35831734
ABSTRACT

BACKGROUND:

Lung cancers can recur locally due to inadequate resection margins. Achieving adequate margin distances is challenging in pulmonary ground glass opacities (GGOs) because they are not easily palpable. To improve margin assessment during resection of GGOs, we propose a novel technique, three-dimensional near-infrared specimen mapping (3D-NSM).

METHODS:

Twenty patients with a cT1 GGO were enrolled and received a fluorescent tracer preoperatively. After resection, specimens underwent 3D-NSM in the operating room. Margins were graded as positive or negative based upon fluorescence at the staple line. Images were analyzed using ImageJ to quantify the distance from the tumor edge to the nearest staple line. This margin distance calculated by 3D-NSM was compared to the margin distance reported on final pathology several days postoperatively.

RESULTS:

3D-NSM identified 20/20 GGOs with no false positive or false negative diagnoses. Mean fluorescence intensity for lesions was 110.92 arbitrary units (A.U.) (IQR 77.77-122.03 A.U.) compared to 23.68 A.U. (IQR 19.60-27.06 A.U.) for background lung parenchyma (p < 0.0001). There were 4 tumor-positive or close margins in the study cohort, and all 4 (100%) were identified by 3D-NSM. 3D-NSM margin distances were nearly identical to margin distances reported on final pathology (R2 = 0.9362). 3D-NSM slightly under-predicted margin distance, and the median difference in margins was 1.9 mm (IQR 0.5-4.3 mm).

CONCLUSIONS:

3D-NSM rapidly localizes GGOs by fluorescence and detects tumor-positive or close surgical margins. 3D-NSM can accurately quantify the resection margin distance as compared to formal pathology, which allows surgeons to rapidly determine whether sublobar resection margin distances are adequate.