Your browser doesn't support javascript.
loading
Comparing the accuracy of three-dimensional mini-optical portable navigation and accelerometer-based portable navigation system for acetabular cup placement during total hip arthroplasty.
Hayashi, Shinya; Kuroda, Yuichi; Nakano, Naoki; Matsumoto, Tomoyuki; Kamenaga, Tomoyuki; Maeda, Toshihisa; Kuroda, Ryosuke.
  • Hayashi S; Department of Orthopaedic Surgery, Kobe University Graduate School of Medicine, 7-5-1 Kusunoki-cho, Chuo-ku, Kobe, 650-0017, Japan. shayashi@med.kobe-u.ac.jp.
  • Kuroda Y; Department of Orthopaedic Surgery, Kobe University Graduate School of Medicine, 7-5-1 Kusunoki-cho, Chuo-ku, Kobe, 650-0017, Japan.
  • Nakano N; Department of Orthopaedic Surgery, Kobe University Graduate School of Medicine, 7-5-1 Kusunoki-cho, Chuo-ku, Kobe, 650-0017, Japan.
  • Matsumoto T; Department of Orthopaedic Surgery, Kobe University Graduate School of Medicine, 7-5-1 Kusunoki-cho, Chuo-ku, Kobe, 650-0017, Japan.
  • Kamenaga T; Department of Orthopaedic Surgery, Kobe University Graduate School of Medicine, 7-5-1 Kusunoki-cho, Chuo-ku, Kobe, 650-0017, Japan.
  • Maeda T; Department of Orthopaedic Surgery, Kobe University Graduate School of Medicine, 7-5-1 Kusunoki-cho, Chuo-ku, Kobe, 650-0017, Japan.
  • Kuroda R; Department of Orthopaedic Surgery, Kobe University Graduate School of Medicine, 7-5-1 Kusunoki-cho, Chuo-ku, Kobe, 650-0017, Japan.
Arch Orthop Trauma Surg ; 143(7): 3759-3766, 2023 Jul.
Article en En | MEDLINE | ID: mdl-36001170
ABSTRACT

INTRODUCTION:

This study compared the accuracy of three dimensional (3D) mini-optical navigation and accelerometer-based portable navigation systems for cup positioning during a total hip arthroplasty (THA) in the supine position. MATERIALS AND

METHODS:

This retrospective cohort study assessed data for 77 hips using 3D mini-optical navigation (n = 37) and accelerometer-based portable navigation (n = 40). The patients underwent THA through the mini-anterolateral approach in the supine position using a portable navigation system. We assessed the preoperative target angles, recorded intraoperative cup angles, postoperative CT imaging angles, cup angle measurement errors, and other clinical parameters.

RESULTS:

The mean absolute differences in radiographic inclination were similar between 3D mini-optical navigation and accelerometer-based portable navigation systems during THA in the supine position (2.8° ± 1.7° vs 2.8° ± 1.9°, p = 0.637). The mean absolute differences in radiographic anteversion were also similar (2.6° ± 2.3° vs 2.5° ± 1.9°, p = 0.737). Cup malalignment (absolute difference of inclination or anteversion between postoperative CT and preoperative target angle of > 5°) was significantly associated with body mass index (BMI) in accelerometer-based portable navigation but not in 3D mini-optical navigation.

CONCLUSIONS:

This is the first study to compare the accuracy of cup positioning between 3D mini-optical and accelerometer-based navigations in THA in the supine position. Both portable navigation systems accurately identified the orientation of cup placement. The accuracy of 3D mini-optical navigation is not affected by high BMI and may be preferred over other options in such patients.