Implementation of a portable diagnostic system for Thomson scattering measurements on an electrothermal arc source.
Rev Sci Instrum
; 93(11): 113526, 2022 Nov 01.
Article
en En
| MEDLINE
| ID: mdl-36461429
To fulfill the increasing needs of diagnostic support for researchers in plasma technology, a portable diagnostic package (PDP) equipped for both laser Thomson scattering (TS) and optical emission spectroscopy has been designed and constructed at Oak Ridge National Laboratory (ORNL), aiming to measure the temperature and number density of electrons and temperatures of ions in plasma devices. The PDP has been initially implemented on a high density and low temperature electrothermal arc source (ET-arc) at ORNL to test its TS capability. TS from the plasmas in the ET-arc has been obtained using the PDP. The electron temperature and number density were determined from TS spectra. These results were then compared to measurements from previous studies on the ET-arc. The TS diagnostic measured 0.8 ± 0.1, 1.3 ± 0.2, and 0.7 ± 0.1 eV and (4.4 ± 0.5) × 1021, (5.9 ± 0.7) × 1021, and (4.3 ± 0.5) x 1021 m-3, respectively, from three lines of sight that transect the plasma column.
Texto completo:
1
Banco de datos:
MEDLINE
Tipo de estudio:
Diagnostic_studies
Idioma:
En
Año:
2022
Tipo del documento:
Article