Radiofrequency antenna concepts for human cardiac MR at 14.0 T.
MAGMA
; 36(2): 257-277, 2023 Apr.
Article
en En
| MEDLINE
| ID: mdl-36920549
OBJECTIVE: To examine the feasibility of human cardiac MR (CMR) at 14.0 T using high-density radiofrequency (RF) dipole transceiver arrays in conjunction with static and dynamic parallel transmission (pTx). MATERIALS AND METHODS: RF arrays comprised of self-grounded bow-tie (SGBT) antennas, bow-tie (BT) antennas, or fractionated dipole (FD) antennas were used in this simulation study. Static and dynamic pTx were applied to enhance transmission field (B1+) uniformity and efficiency in the heart of the human voxel model. B1+ distribution and maximum specific absorption rate averaged over 10 g tissue (SAR10g) were examined at 7.0 T and 14.0 T. RESULTS: At 14.0 T static pTx revealed a minimum B1+ROI efficiency of 0.91 µT/âkW (SGBT), 0.73 µT/âkW (BT), and 0.56 µT/âkW (FD) and maximum SAR10g of 4.24 W/kg, 1.45 W/kg, and 2.04 W/kg. Dynamic pTx with 8 kT points indicate a balance between B1+ROI homogeneity (coefficient of variation < 14%) and efficiency (minimum B1+ROI > 1.11 µT/âkW) at 14.0 T with a maximum SAR10g < 5.25 W/kg. DISCUSSION: MRI of the human heart at 14.0 T is feasible from an electrodynamic and theoretical standpoint, provided that multi-channel high-density antennas are arranged accordingly. These findings provide a technical foundation for further explorations into CMR at 14.0 T.
Palabras clave
Texto completo:
1
Banco de datos:
MEDLINE
Asunto principal:
Imagen por Resonancia Magnética
/
Corazón
Tipo de estudio:
Prognostic_studies
Límite:
Humans
Idioma:
En
Año:
2023
Tipo del documento:
Article