Deep learning for automated left ventricular outflow tract diameter measurements in 2D echocardiography.
Zha, Sigurd Zijun; Rogstadkjernet, Magnus; Klæboe, Lars Gunnar; Skulstad, Helge; Singstad, Bjørn-Jostein; Gilbert, Andrew; Edvardsen, Thor; Samset, Eigil; Brekke, Pål Haugar.
  • Zha SZ; University of Oslo, Oslo, Norway. sigzha@gmail.com.
  • Rogstadkjernet M; University of Oslo, Oslo, Norway.
  • Klæboe LG; Oslo University Hospital, Rikshospitalet, Oslo, Norway.
  • Skulstad H; University of Oslo, Oslo, Norway.
  • Singstad BJ; Oslo University Hospital, Rikshospitalet, Oslo, Norway.
  • Gilbert A; Oslo University Hospital, Rikshospitalet, Oslo, Norway.
  • Edvardsen T; GE HealthCare, Oslo, Norway.
  • Samset E; University of Oslo, Oslo, Norway.
  • Brekke PH; Oslo University Hospital, Rikshospitalet, Oslo, Norway.
Cardiovasc Ultrasound ; 21(1): 19, 2023 Oct 13.
Article en En | MEDLINE | ID: mdl-37833731
ABSTRACT

BACKGROUND:

Measurement of the left ventricular outflow tract diameter (LVOTd) in echocardiography is a common source of error when used to calculate the stroke volume. The aim of this study is to assess whether a deep learning (DL) model, trained on a clinical echocardiographic dataset, can perform automatic LVOTd measurements on par with expert cardiologists.

METHODS:

Data consisted of 649 consecutive transthoracic echocardiographic examinations of patients with coronary artery disease admitted to a university hospital. 1304 LVOTd measurements in the parasternal long axis (PLAX) and zoomed parasternal long axis views (ZPLAX) were collected, with each patient having 1-6 measurements per examination. Data quality control was performed by an expert cardiologist, and spatial geometry data was preserved for each LVOTd measurement to convert DL predictions into metric units. A convolutional neural network based on the U-Net was used as the DL model.

RESULTS:

The mean absolute LVOTd error was 1.04 (95% confidence interval [CI] 0.90-1.19) mm for DL predictions on the test set. The mean relative LVOTd errors across all data subgroups ranged from 3.8 to 5.1% for the test set. Generally, the DL model had superior performance on the ZPLAX view compared to the PLAX view. DL model precision for patients with repeated LVOTd measurements had a mean coefficient of variation of 2.2 (95% CI 1.6-2.7) %, which was comparable to the clinicians for the test set.

CONCLUSION:

DL for automatic LVOTd measurements in PLAX and ZPLAX views is feasible when trained on a limited clinical dataset. While the DL predicted LVOTd measurements were within the expected range of clinical inter-observer variability, the robustness of the DL model requires validation on independent datasets. Future experiments using temporal information and anatomical constraints could improve valvular identification and reduce outliers, which are challenges that must be addressed before clinical utilization.