Effect of decentration on the quality of vision in two aspheric posterior chamber intraocular lenses: A contralateral eye study.
Thakur, Anchal; Adiga, Sushant; Malhotra, Chintan; Sachdeva, Kulbhushan; Singh, Shubhi; Jain, Arun Kumar.
  • Thakur A; Advanced Eye Centre, PGIMER, Chandigarh, India.
Indian J Ophthalmol ; 72(4): 558-564, 2024 Apr 01.
Article en En | MEDLINE | ID: mdl-38189441
ABSTRACT

PURPOSE:

To perform an intraindividual comparison of the quality of vision and the effect of decentration between two aspheric intraocular lenses aspheric balanced curve (ABC) design Vivinex iSert XY1 (Hoya Surgical Optics, Singapore) and anterior aspheric design Tecnis ZCB00 (Abbott Medical Optics, CA).

SETTING:

Tertiary Eye Care Centre.

DESIGN:

Prospective, randomized comparative study using a random number table.

METHODS:

Thirty patients were randomized to the implantation of Vivinex iSert XY1 in one eye and Tecnis ZCB00 in the contralateral eye. Then, 12 weeks postoperatively, a laser ray-tracing aberrometer was used to evaluate the visual Strehl ratio, higher-order aberrations (HOA), decentration of IOL from the visual axis and geometric axis, angle alpha, and angle kappa. Contrast sensitivity was measured using the functional visual analyzer.

RESULTS:

The visual internal Strehl ratio was higher ( P < 0.05) at all pupil sizes and the spherical aberrations values were lower ( P < 0.05) at larger pupil sizes (5 mm and 6 mm) in the Vivinex group. The mean decentration from the visual axis in the Vivinex group was significantly more than that in the Tecnis group ( P < 0.01). With an increase in angle alpha, there was a greater decline in the visual Strehl ratio for Tecnis; however, there was a statistically significant decline at 3 mm pupil size for Tecnis ZCB00 ( P = 0.028). The contrast sensitivity was similar for both IOLs.

CONCLUSION:

In comparison to an anterior aspheric design IOL, the ABC design IOL yielded better quality of vision, neutralized spherical aberrations to a larger extent, and provided a relatively superior quality of vision with decentration.