Your browser doesn't support javascript.
loading
Spectral performance evaluation of a second-generation spectral detector CT.
Liu, Leening P; Shapira, Nadav; Halliburton, Sandra S; Meyer, Sebastian; Perkins, Amy; Litt, Harold I; Kauczor, Hans Ulrich; Leiner, Tim; Stiller, Wolfram; Noël, Peter B.
  • Liu LP; Department of Radiology, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Shapira N; Department of Bioengineering, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Halliburton SS; Department of Radiology, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Meyer S; Philips Healthcare, Orange Village, Ohio, USA.
  • Perkins A; Department of Radiology, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Litt HI; Philips Healthcare, Orange Village, Ohio, USA.
  • Kauczor HU; Department of Radiology, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Leiner T; Diagnostic and Interventional Radiology (DIR), Heidelberg University Hospital, Heidelberg, Germany.
  • Stiller W; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Noël PB; Diagnostic and Interventional Radiology (DIR), Heidelberg University Hospital, Heidelberg, Germany.
J Appl Clin Med Phys ; 25(4): e14300, 2024 Apr.
Article en En | MEDLINE | ID: mdl-38386967
ABSTRACT

PURPOSE:

The aim of this study was to characterize a second-generation wide-detector dual-layer spectral computed tomography (CT) system for material quantification accuracy, acquisition parameter and patient size dependencies, and tissue characterization capabilities.

METHODS:

A phantom with multiple tissue-mimicking and material-specific inserts was scanned with a dual-layer spectral detector CT using different tube voltages, collimation widths, radiation dose levels, and size configurations. Accuracy of iodine density maps and virtual monoenergetic images (MonoE) were investigated. Additionally, differences between conventional and MonoE 70 keV images were calculated to evaluate acquisition parameter and patient size dependencies. To demonstrate material quantification and differentiation, liver-mimicking inserts with adipose and iron were analyzed with a two-base decomposition utilizing MonoE 50 and 150 keV, and root mean square error (RMSE) for adipose and iron content was reported.

RESULTS:

Measured inserts exhibited quantitative accuracy across a wide range of MonoE levels. MonoE 70 keV images demonstrated reduced dependence compared to conventional images for phantom size (1 vs. 27 HU) and acquisition parameters, particularly tube voltage (4 vs. 37 HU). Iodine density quantification was successful with errors ranging from -0.58 to 0.44 mg/mL. Similarly, inserts with different amounts of adipose and iron were differentiated, and the small deviation in values within inserts corresponded to a RMSE of 3.49 ± 1.76% and 1.67 ± 0.84 mg/mL for adipose and iron content, respectively.

CONCLUSION:

The second-generation dual-layer CT enables acquisition of quantitatively accurate spectral data without compromises from differences in patient size and acquisition parameters.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Tomografía Computarizada por Rayos X / Yodo Límite: Humans Idioma: En Año: 2024 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Tomografía Computarizada por Rayos X / Yodo Límite: Humans Idioma: En Año: 2024 Tipo del documento: Article