Effect of eccentric fixation on the steady-state pattern electroretinogram.
Friedel, Evelyn B N; Haldina, Julia; Nickel, Kathrin; Bach, Michael; Tebartz van Elst, Ludger; Heinrich, Sven P.
  • Friedel EBN; Eye Center, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany. evelyn.friedel@uniklinik-freiburg.de.
  • Haldina J; Department of Psychiatry and Psychotherapy, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany. evelyn.friedel@uniklinik-freiburg.de.
  • Nickel K; Faculty of Biology, University of Freiburg, Freiburg, Germany. evelyn.friedel@uniklinik-freiburg.de.
  • Bach M; Eye Center, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
  • Tebartz van Elst L; Department of Psychiatry and Psychotherapy, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
  • Heinrich SP; Eye Center, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
Doc Ophthalmol ; 148(2): 87-95, 2024 Apr.
Article en En | MEDLINE | ID: mdl-38416305
ABSTRACT

PURPOSE:

The steady-state pattern electroretinogram (ssPERG) is used to assess retinal ganglion cell function in a variety of research contexts and diagnostic applications. In certain groups of patients or study participants, stable central fixation of the stimulus is not guaranteed. The present study aimed at assessing the effects of misfixation on the ssPERG response to checkerboard reversal stimuli.

METHODS:

Using two check sizes (0.8° and 15°), we compared ssPERG responses for several amounts of fixation deviation, ranging from 0° to 19° horizontally and from 0° to 14° diagonally. The stimulus area extended to 15° eccentricity, stimulus reversal rate was 15/s.

RESULTS:

Up to around 7° eccentricity, there was no sizable effect of fixation deviation under most conditions. Effects were somewhat larger for nasal than for temporal deviation, in particular for small checks. Diagonal deviation was associated with a response to luminance onset/offset at 7.5 Hz (subharmonic of the reversal rate), most prominently when the interior of a large check was fixated.

CONCLUSION:

Generally, moderate inaccuracies of fixation do not have a sizable effect on ssPERG amplitude. However, with large checks, the luminance response has to be considered.