High throughput multidimensional liquid chromatography approach for online protein removal and characterization of polysorbates and poloxamer in monoclonal antibody formulations.
Zegota, Maksymilian M; Schuster, Georg; De Pra, Mauro; Müllner, Tibor; Menzen, Tim; Steiner, Frank; Hawe, Andrea.
  • Zegota MM; Coriolis Pharma Research, Fraunhoferstraße 18B, 82152 Martinsried, Germany. Electronic address: maksymilian.zegota@coriolis-pharma.com.
  • Schuster G; Coriolis Pharma Research, Fraunhoferstraße 18B, 82152 Martinsried, Germany.
  • De Pra M; Thermo Fisher Scientific, Dornierstraße 4, 82110 Germering, Germany.
  • Müllner T; Thermo Fisher Scientific, Dornierstraße 4, 82110 Germering, Germany.
  • Menzen T; Coriolis Pharma Research, Fraunhoferstraße 18B, 82152 Martinsried, Germany.
  • Steiner F; Thermo Fisher Scientific, Dornierstraße 4, 82110 Germering, Germany.
  • Hawe A; Coriolis Pharma Research, Fraunhoferstraße 18B, 82152 Martinsried, Germany.
J Chromatogr A ; 1720: 464777, 2024 Apr 12.
Article en En | MEDLINE | ID: mdl-38432108
ABSTRACT
The majority of commercially available monoclonal antibody (mAb) formulations are stabilized with one of three non-ionic surfactants polysorbate 20 (PS20), polysorbate 80 (PS80), or poloxamer 188 (P188). All three surfactants are susceptible to degradation, which can result in functionality loss and subsequent protein aggregation or free fatty acid particle formation. Consequently, quantitative, and qualitative analysis of surfactants is an integral part of formulation development, stability, and batch release testing. Due to the heterogeneous nature of both polysorbates and poloxamer, online isolation of all the compounds from the protein and other excipients that may disturb the subsequent liquid chromatography with charged aerosol detection (LC-CAD) analysis poses a challenge. Herein, we present an analytical method employing LC-CAD, utilizing a combination of anion and cation exchange columns to completely remove proteins online before infusing the isolated surfactant onto a reversed-phase column. The method allows high throughput analysis of polysorbates within 8 minutes and poloxamer 188 within 12 minutes, providing a separation of the surfactant species of polysorbates (unesterified species, lower esters, and higher esters) and poloxamer 188 (early eluters and main species). Accuracy and precision assessed according to the International Council for harmonisation (ICH) guideline were 96 - 109 % and ≤1 % relative standard deviation respectively for all three surfactants in samples containing up to 110 mg/mL mAb. Subsequently, the method was effectively applied to quantify polysorbate 20 and polysorbate 80 in nine commercial drug products with mAb concentration of up to 180 mg/mL.