Assessing the influence of parameters on tissue welding in small bowel end-to-end anastomosis in vitro and in vivo.
Zhu, Caihui; Na, Yuyan; Cheng, Xiujun; Tao, Xiaonan; Xie, Pengyao; Chen, Lei; Zhao, Hui; Qiu, Jian; Gu, Xiaodong; Xiang, Jianbin; Liu, Kefu.
  • Zhu C; School of Information Science and Technology, Fudan University, Shanghai, China.
  • Na Y; Department of Sports Medicine, Huashan Hospital, Fudan University, Shanghai, China.
  • Cheng X; Department of Dermatology, Huashan Hospital, Fudan University, Shanghai, China.
  • Tao X; School of Information Science and Technology, Fudan University, Shanghai, China.
  • Xie P; School of Information Science and Technology, Fudan University, Shanghai, China.
  • Chen L; School of Information Science and Technology, Fudan University, Shanghai, China.
  • Zhao H; School of Information Science and Technology, Fudan University, Shanghai, China.
  • Qiu J; School of Information Science and Technology, Fudan University, Shanghai, China.
  • Gu X; Department of General Surgery, Huashan Hospital, Fudan University, Shanghai, China.
  • Xiang J; Department of General Surgery, Huashan Hospital, Fudan University, Shanghai, China.
  • Liu K; School of Information Science and Technology, Fudan University, Shanghai, China. kfliu@fudan.edu.cn.
Surg Endosc ; 38(6): 3126-3137, 2024 Jun.
Article en En | MEDLINE | ID: mdl-38622226
ABSTRACT

BACKGROUND:

The use of high-frequency electric welding technology for intestinal end-to-end anastomosis holds significant promise. Past studies have focused on in vitro, and the safety and efficacy of this technology is uncertain, severely limiting the clinical application of this technology. This study investigates the impact of compression pressure, energy dosage, and duration on anastomotic quality using a homemade anastomosis device in both in vitro and in vivo settings.

METHODS:

Two hundred eighty intestines and 5 experimental pigs were used for in vitro and in vivo experiments, respectively. The in vitro experiments were conducted to study the effects of initial pressure (50-400 kpa), voltage (40-60 V), and time (10-20 s) on burst pressure, breaking strength, thermal damage, and histopathological microstructure of the anastomosis. Optimal parameters were then inlaid into a homemade anastomosis and used for in vivo experiments to study the postoperative porcine survival rate and the pathological structure of the tissues at the anastomosis and the characteristics of the collagen fibers.

RESULTS:

The anastomotic strength was highest when the compression pressure was 250 kPa, the voltage was 60 V, and the time was 15 s. The degree of thermal damage to the surrounding tissues was the lowest. The experimental pigs had no adverse reactions after the operation, and the survival rate was 100%. 30 days after the operation, the surgical site healed well, and the tissues at the anastomosis changed from immediate adhesions to permanent connections.

CONCLUSION:

High-frequency electric welding technology has a certain degree of safety and effectiveness. It has the potential to replace the stapler anastomosis in future and become the next generation of new anastomosis device.